Results:
 +
{| border="0" style="text-align:center;" width="100%"
 +
|+
 +
|-
 +
| valign="top"|
 +
[[File:benchm_transverse_plot_lid_cloth.jpg|thumb|center|340px|Measurements of transverse transmittance of BTCM crystal with lid and cloth covering the spectrometer's sample compartment.]]
 +
| valign="top"|
 +
[[File:benchm_transverse_diff_lid_cloth.jpg|thumb|center|340px|Difference between these two measurements.]]
 +
|}

Revision as of 12:15, 9 November 2015

⇐ Back to Neutral Particle Spectrometer
⇐ Back to the Main_Page

Spectrometer benchmark

A set of tests were performed in order to understand the repetitivity, reprodutibility, and the effects of some components of the spectrometer in the transmittance measurements of PWO crystals with the Lambda 750 UV/VIS/NIR spectrometer at CUA.

Two types of measurements are planned to be performed with this spectrometers to quantify the crystals transmittance: longitudinal measurement and transverse scan along crystal length. For each of these, a set of benchmark tests were done as described in the following subsections.

Transverse Transmittance

Cloth to cover the spectrometer

This test allowed us to make sure there is no effect on the transmittance measurements when working with a cloth covering the sample compartment instead the spectrometer's lid. Using the cloth will be necessary since the 20 cm crystals can only be fully scanned in the spectrometer if the lid is removed, since the sample compartment is not big enough.

Measurements:

  • 1 - Measure transmittance of a 20-cm BTCP crystal installed at the center of the compartment with the compartment cover installed and closed.
  • 2 - Measure transmittance of 20-cm BTCP crystal installed at the center of the compartment without the compartment cover and a black cloth covering the setup. Do not move the crystal between point (1) and this measurement. This test is done without the diagonal slider. Compare the result with that of the previous point. If there is significant noise in the data, check the positioning of the black cloth.

Pictures:

File:Benchm transverse lid.jpg
Spectrometer with lid.
Spectrometer with cloth.

Results:

File:Benchm transverse plot lid cloth.jpg
Measurements of transverse transmittance of BTCM crystal with lid and cloth covering the spectrometer's sample compartment.
File:Benchm transverse diff lid cloth.jpg
Difference between these two measurements.