Difference between revisions of "MainPage:Nuclear:NPS:PWO:CrystalLOG:SICCA15"

From cua_phy
Jump to navigation Jump to search
 
(6 intermediate revisions by 2 users not shown)
Line 24: Line 24:
 
| 01/14/2016 || Transverse Transmittance || Spectral measurement.
 
| 01/14/2016 || Transverse Transmittance || Spectral measurement.
 
|-
 
|-
 +
| 06/01/2016 || Longitudinal Transmittance || Spectral measurement, Lambda 950 (integrating sphere). Wrapped crystal.
 +
|-
 +
| 06/02/2016 || Longitudinal Transmittance || Repeating measurement, to make sure alignment and other effects are indeed negligible in our setup.
 +
|-
 +
| 06/03/2016 || Thermal annealing || Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour.
 +
|-
 +
| 06/06/2016 || Transverse Transmittance || Spectral measurement with Lambda 750.
 +
|-
 +
| 06/06/2016 || Transverse Transmittance || Spectral measurement with Lambda 950 - integrated sphere (may not be at exactly the same position).
 +
|-
 +
| 06/06/2016 || Crystal wrapped || 5 layers of teflon tape and 2 layers of black electrical tape.
 +
|-
 +
| 06/06/2016 || Longitudinal Transmittance || Spectral measurement with Lambda 950 - integrated sphere.
 +
|}
 +
 +
== Transmittance measurements ==
 +
 +
=== Longitudinal transmittance ===
 +
 +
{| border="0" style="text-align:center;" width="100%"
 +
|-
 +
| valign="top"|
 +
[[File:SICCA 15 LT BI.png|thumb|center|340px|Longitudinal transmittance measured at Lambda 750 (no integrating sphere). Crystal as arrived at CUA (probably irradiated at Caltech).]]
 +
| valign="top"|
 +
[[File:SICCA15_LT_sphere_anneal.png|thumb|center|340px|Longitudinal transmittance measured at Lambda 950 (with integrating sphere).]]
 
|}
 
|}
  
Line 31: Line 56:
 
|-
 
|-
 
| valign="top"|
 
| valign="top"|
[[File:Sic15_1.png|thumb|center|340px|xxxx.]]
+
[[File:Sic15_1.png|thumb|center|340px|Close-up view of crack on edge..]]
 
| valign="top"|
 
| valign="top"|
[[File:Sic15_2.png|thumb|center|340px|xxxx.]]
+
[[File:Sic15_2.png|thumb|center|340px|View of crystal and crack on edge.]]
 
|-
 
|-
 
|}
 
|}
Line 39: Line 64:
 
|-
 
|-
 
| valign="top"|
 
| valign="top"|
[[File:Sic15_3|thumb|center|340px|Carving of "PWO SIC 15" clearly shown on crystal.]]
+
[[File:Sic15_3.png|thumb|center|340px|Carving of "PWO SIC 15" clearly shown on crystal, along with an even closer up view of crack on the edge.]]
 
|-
 
|-
 
|}
 
|}

Latest revision as of 15:46, 6 June 2016

⇐ Back to Crystal LOG
⇐ Back to Neutral Particle Spectrometer
⇐ Back to the Main_Page

Time LOG

SICCA 15
Date Description Comments
01/10/2016 Pictures Documentation as arrived at CUA (received from Caltech). Name of the crystal was observed scratched on the face. Pictures 01/10/2016.
01/14/2016 Long. Transmittance Crystal unwrapped.
01/14/2016 Transverse Transmittance Spectral measurement.
06/01/2016 Longitudinal Transmittance Spectral measurement, Lambda 950 (integrating sphere). Wrapped crystal.
06/02/2016 Longitudinal Transmittance Repeating measurement, to make sure alignment and other effects are indeed negligible in our setup.
06/03/2016 Thermal annealing Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour.
06/06/2016 Transverse Transmittance Spectral measurement with Lambda 750.
06/06/2016 Transverse Transmittance Spectral measurement with Lambda 950 - integrated sphere (may not be at exactly the same position).
06/06/2016 Crystal wrapped 5 layers of teflon tape and 2 layers of black electrical tape.
06/06/2016 Longitudinal Transmittance Spectral measurement with Lambda 950 - integrated sphere.

Transmittance measurements

Longitudinal transmittance

Longitudinal transmittance measured at Lambda 750 (no integrating sphere). Crystal as arrived at CUA (probably irradiated at Caltech).
Longitudinal transmittance measured at Lambda 950 (with integrating sphere).

Pictures at 01/10/2016

Close-up view of crack on edge..
View of crystal and crack on edge.
Carving of "PWO SIC 15" clearly shown on crystal, along with an even closer up view of crack on the edge.