Difference between revisions of "MainPage:Nuclear:NPS:PWO:CrystalLOG:SICCAJ19"
Jump to navigation
Jump to search
Line 71: | Line 71: | ||
| valign="top"| | | valign="top"| | ||
[[File:J19angle.png|thumb|center|340px|Ideal crystal placement would be 20cm side parallel to light beam. Reasonable placement errors would be 1.3° and 2.3° off of this. At these angles the beam still appeared to fall within the crystla face under the "Align" function]] | [[File:J19angle.png|thumb|center|340px|Ideal crystal placement would be 20cm side parallel to light beam. Reasonable placement errors would be 1.3° and 2.3° off of this. At these angles the beam still appeared to fall within the crystla face under the "Align" function]] | ||
+ | |} | ||
== Light Yield == | == Light Yield == |
Revision as of 15:02, 8 August 2016
⇐ Back to Crystal LOG |
⇐ Back to Neutral Particle Spectrometer |
⇐ Back to the Main_Page |
Time LOG
Date | Description | Comments |
---|---|---|
12/14/2015 | Pictures | Documentation of crystal's conditions as it arrived at CUA. |
12/15/2015 | Thermal annealing | Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. |
12/17/2015 | Transverse Transmittance | Scan at 360, 420, and 620nm. |
12/18/2015 | Irradiation | 160 keV photons at 6mA, for 10min. |
12/18/2015 | Transverse Transmittance | Scan at 360, 420, and 620nm after 30 min of irradiation. |
01/06/2016 | LY scan | At 100ns, as described in run 3097. |
03/04/2016 | Transverse Transmittance | At the center of the crystal, every nm from 200nm until 800nm. Sample: 1314 |
03/04/2016 | Dimensions | Crystal's dimensions measured. Length and six points of width (3 in each face). |
05/23/2016 | Transverse Transmittance Scan | Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s. |
07/07/2016 | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Tested transmittance dependence on which 20x2 face of crystal was face down on the Labjack. |
07/07/2016 | Longitudinal Transmittance Scan | Wavelengths: 200-800nm. Measured effect of off center crystal placement on transmittance. +1.3°, +2.3°, -1.3°, -2.3°. |
07/07/2016 | Longitudinal Transmittance Scan | Wavelengths: 200-800nm. Measured effect of iris diameter on transmittance. 8.0mm, 11.5mm(used), 15mm, and 17mm. |
07/11/2016 | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Measured effect of which face of crystal faced light beam in spectrometer. "flip" of crystal. |
07/07/2016 | Longitudinal Transmittance Scan | Wavelengths: 200-800nm. Remeasured effect of off center crystal placement on transmittance BUT with removal of all tape/residue of 2x2 crystal face. +1.3°, +2.3°, -1.3°, -2.3°. |
07/26/2016 | Light Yield | Run:3248 3 layers of teflon tape and 1 layer of elecrtical. ~50,000 incidents recorded at ~18 degrees Celsius. |
07/26/2016 | Transverse Transmittance | 200nm-800nm. 10cm down crystal. |
Pictures at 12/14/2015
Transmittance Uncertainty 07/2016
Light Yield
|}