| 11/30/2016 || Salina || Light Yield Scan || Run 3434: ~1.4M events; Re-run of 10 cm crytur crystal wrapped with ~5 layers of teflon tape and 2 layers of black electrical tape. T=18°C. '''LY = 9.67 ± 0.53 pe/MeV''' (see pictures below).  
 
| 11/30/2016 || Salina || Light Yield Scan || Run 3434: ~1.4M events; Re-run of 10 cm crytur crystal wrapped with ~5 layers of teflon tape and 2 layers of black electrical tape. T=18°C. '''LY = 9.67 ± 0.53 pe/MeV''' (see pictures below).  
 +
|-
 +
| 12/02/2016 || Rich  || Longitudinal Transmittance || Before annealing.
 
|}
 
|}
  

Revision as of 17:28, 2 December 2016

⇐ Back to Crystal LOG
⇐ Back to Neutral Particle Spectrometer
⇐ Back to the Main_Page

Time LOG

CRYTUR 01
Date Experimenter(s) Description Comments
xx/xx/2015 Marco Wrapped Preparing for light yield measurement. Approximately 5 layers of teflon tape, and 2 layers of black tape.
xx/xx/2015 Marco LY scan ...
12/14/2015 Marco unwrapped Preparing for annealing.
12/14/2015 Marco Thermal annealing Starting 11:00AM. Ramp up at 18oC/hour until 200oC, stay at 200oC for 10 hours, ramp down at -18oC/hour.
12/17/2015 Marco Longitudinal Transmittance After annealing.
07/15/2016 Summer Students Longitudinal Transmittance Wavelength:200-800nm.
08/11/2016 Summer Students LY scan RUN:3411. Motor was lowered to approximately 5 cm down crystal. ~18oC. LY: 11.986 phe/MeV-
11/18/2016 Rich Light Yield Scan Run 3422: ~100 events; 10 cm crytur crystal wrapped with ~4 layers of teflon tape and 2 layers of black electrical tape. T=18°C. LY = 9.3978 pe/Mev (see pictures below). Not enough runs? Note: the error bars used were not the 0.5 standard used previously, but instead the calculated uncertainty for each run.
11/29/2016 Salina Light Yield Scan Run 3432: ~100K events; 10 cm crytur crystal wrapped with ~5 layers of teflon tape and 2 layers of black electrical tape. T=18°C. LY = 8.23 ± 1.08 pe/MeV. Not enough statistics? Need to re-run with more stats (see pictures below)
11/30/2016 Salina Light Yield Scan Run 3434: ~1.4M events; Re-run of 10 cm crytur crystal wrapped with ~5 layers of teflon tape and 2 layers of black electrical tape. T=18°C. LY = 9.67 ± 0.53 pe/MeV (see pictures below).
12/02/2016 Rich Longitudinal Transmittance Before annealing.

Transmittance measurements

Longitudinal transmittance measured at Lambda 750 (no integrating sphere).
Longitudinal transmittance measured at Lambda 750 (no integrating sphere).

Pictures at 12/14/2015

CRYTUR shown inside oven for thermal annealing.

Light Yield

Run 3422 taken on 11/18/2016: Fit of histogram with ~100K events.
Run 3422 taken on 11/18/2016: LY as a function of position. Error bars based off calculatd uncertainty.
Run 3432 taken on 11/29/2016: Fit of histogram with ~100K events.
Run 3432 taken on 11/29/2016: LY as a function of position.
Run 3434 taken on 11/29/2016: Fit of histogram with ~1.4M events.
Run 3434 taken on 11/29/2016: LY as a function of position.

Additional Data/Info

Graph showing CRYTUR crystal longitudinal measurements before annealing (red) and after annealing (green).
Graph showing the difference of the CRYTUR crystal longitudinal measurements before and after annealing.