Difference between revisions of "MainPage:Nuclear:NPS:PWO:CrystalLOG:SICCAJ18"
Jump to navigation
Jump to search
(23 intermediate revisions by 4 users not shown) | |||
Line 15: | Line 15: | ||
! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:10%;" | Date | ! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:10%;" | Date | ||
+ | ! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:15%;" | Experimenter(s) | ||
! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:15%;" | Description | ! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:15%;" | Description | ||
! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:75%;" | Comments | ! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:75%;" | Comments | ||
|- | |- | ||
− | | 12/14/2015 || Pictures || Documentation of crystal's conditions as it arrived at CUA. | + | | 12/14/2015 || Marco || Pictures || Documentation of crystal's conditions as it arrived at CUA. |
|- | |- | ||
− | | 12/15/2015 || Thermal annealing || Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. | + | | 12/15/2015 || Marco || Thermal annealing || Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. |
|- | |- | ||
− | | 12/17/2015 || Long. Transmittance || Note: crystal NOT wrapped. | + | | 12/17/2015 || Marco || Long. Transmittance || Note: crystal NOT wrapped. |
|- | |- | ||
− | | 12/17/2015 || Transverse Transmittance || Scan at 360, 420, and 620nm. | + | | 12/17/2015 || Marco || Transverse Transmittance || Scan at 360, 420, and 620nm. |
|- | |- | ||
− | | 12/18/2015 || Irradiation || 160 keV photons at 6mA, for 10min. | + | | 12/18/2015 || Marco || Irradiation || 160 keV photons at 6mA, for 10min. |
|- | |- | ||
− | | 12/18/2015 || Transverse Transmittance || Scan at 360, 420, and 620nm after 30 min of irradiation. | + | | 12/18/2015 || Marco || Transverse Transmittance || Scan at 360, 420, and 620nm after 30 min of irradiation. |
|- | |- | ||
− | | 01/05/2016 || LY scan || At 100ns, as described in run 3047. | + | | 01/05/2016 || Marco || LY scan || At 100ns, as described in run 3047. |
|- | |- | ||
− | | 03/04/2016 || Transverse Transmittance || At the center of the crystal, every nm from 200nm until 800nm. Sample: 1315 | + | | 03/04/2016 || Marco || Transverse Transmittance || At the center of the crystal, every nm from 200nm until 800nm. Sample: 1315 |
|- | |- | ||
− | | 03/04/2016 || Dimensions || Crystal's dimensions measured. Length and six points of width (3 in each face). | + | | 03/04/2016 || Marco || Dimensions || Crystal's dimensions measured. Length and six points of width (3 in each face). |
|- | |- | ||
− | | 05/23/2016 || Transverse Transmittance Scan || Wavelengths: | + | | 05/23/2016 || Marco || Transverse Transmittance Scan || Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s. |
|- | |- | ||
− | | 07/ | + | | 07/06/2016 || Summer Students || Longitudinal Transmittance Scan || Wavelengths: 200-800 nm. Measured effect of off-center placement of crystal on transmittance. +1.3°, +2.3°, -1.3°, -2.3. |
+ | |- | ||
+ | | 07/06/2016 || Summer Students || Longitudinal Transmittance Scan || Wavelengths: 200-800 nm. Measured light beam entering both end face of crystal in spectrometer. "flip" of crystal. | ||
+ | |- | ||
+ | | 07/07/2016 || Summer Students || Longitudinal Transmittance Scan || Wavelengths: 200-800 nm. Measured effect of iris diameter on transmittace. 8mm, 11.5mm (used), 15mm, 17mm. | ||
+ | |- | ||
+ | | 07/14/2016 || Summer Students || Longitudinal Transmittance Scan || Wavelengths: 200-800 nm. Remeasured effect of off-center placement of crystal on transmittance BUT with removal of all residue/tape off 2x2 crystal face. +1.3°, +2.3°, -1.3°, -2.3. | ||
+ | |- | ||
+ | | 07/21/2016 || Summer Students || Light Yield (unmeasured) || Crystal accident , crack on 2x2 face of crystal. | ||
+ | |- | ||
+ | | 07/28/2016 || Summer Students || Transverse Transmittance || 200-800nm wavelengths. 10cm down crystal. | ||
+ | |- | ||
+ | | 12/20/2016 || TH || Light Yield || run 3507: '''11.28 pe/MeV''', note: not a good fit to SEP, low stat run. | ||
+ | |- | ||
+ | | 07/28/2016 || Summer Students || Longitudinal Transmittance || 200-800nm wavelengths. Sampling period: 0.6s, moving motor at 2mm/s. | ||
+ | |- | ||
+ | | 07/07/2017 || Summer Students || Longitudinal Transmittance || 200-800nm wavelengths. Sampling period: 0.6s, moving motor at 2mm/s. | ||
+ | |- | ||
+ | | 07/10/2017 || Summer Students || Light Yield || Run 3691. 100ns gw. ~50,000 incidents ~3 layers of teflon tape ~1 layer of electrical. LY=10.01 pe/MeV. | ||
|} | |} | ||
Line 47: | Line 66: | ||
| valign="top"| | | valign="top"| | ||
[[File:SICCA J18 LT.png|thumb|center|340px|Longitudinal transmittance measured at Lambda 750 (no integrating sphere).]] | [[File:SICCA J18 LT.png|thumb|center|340px|Longitudinal transmittance measured at Lambda 750 (no integrating sphere).]] | ||
+ | | valign="top"| | ||
+ | |- | ||
+ | | valign="top"| | ||
+ | [[File:J18longtrans.png|thumb|center|340px|Longitudinal transmittance measured at Lambda 950 (integrating sphere).]] | ||
| valign="top"| | | valign="top"| | ||
|} | |} | ||
Line 58: | Line 81: | ||
| valign="top"| | | valign="top"| | ||
[[File:SICCAJ18.JPG|thumb|center|340px|xxxx.]] | [[File:SICCAJ18.JPG|thumb|center|340px|xxxx.]] | ||
+ | | valign="top"| | ||
+ | |- | ||
+ | [[File:J18crack.JPG|thumb|left|340px|On July 21 when crystal was put in to test light yield (light yield was not taken).]] | ||
+ | |} | ||
+ | |||
+ | == Transmittance Uncertainty == | ||
+ | |||
+ | {| border="0" style="text-align:center;" width="100%" | ||
+ | |- | ||
+ | | valign="top"| | ||
+ | [[File:J18iris.png|thumb|center|340px|Iris Diameter vs Transmittance]] | ||
+ | | valign="top"| | ||
+ | [[File:J18degree.png|thumb|center|340px|Ideal crystal placement would be 20cm side parallel to light beam. Reasonable placement errors would be 1.3° and 2.3° off of this. At these angles the beam still appeared to fall within the crystla face under the "Align" function]] | ||
+ | |- | ||
+ | | valign="top"| | ||
+ | [[File:J18flip.png|thumb|center|340px|Internal Symmetry of Crystal. "Flipping" the crystal by reversing which end-face the light beam hits first.]] | ||
|} | |} |
Latest revision as of 11:43, 11 July 2017
⇐ Back to Crystal LOG |
⇐ Back to Neutral Particle Spectrometer |
⇐ Back to the Main_Page |
Time LOG
Date | Experimenter(s) | Description | Comments |
---|---|---|---|
12/14/2015 | Marco | Pictures | Documentation of crystal's conditions as it arrived at CUA. |
12/15/2015 | Marco | Thermal annealing | Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. |
12/17/2015 | Marco | Long. Transmittance | Note: crystal NOT wrapped. |
12/17/2015 | Marco | Transverse Transmittance | Scan at 360, 420, and 620nm. |
12/18/2015 | Marco | Irradiation | 160 keV photons at 6mA, for 10min. |
12/18/2015 | Marco | Transverse Transmittance | Scan at 360, 420, and 620nm after 30 min of irradiation. |
01/05/2016 | Marco | LY scan | At 100ns, as described in run 3047. |
03/04/2016 | Marco | Transverse Transmittance | At the center of the crystal, every nm from 200nm until 800nm. Sample: 1315 |
03/04/2016 | Marco | Dimensions | Crystal's dimensions measured. Length and six points of width (3 in each face). |
05/23/2016 | Marco | Transverse Transmittance Scan | Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s. |
07/06/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Measured effect of off-center placement of crystal on transmittance. +1.3°, +2.3°, -1.3°, -2.3. |
07/06/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Measured light beam entering both end face of crystal in spectrometer. "flip" of crystal. |
07/07/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Measured effect of iris diameter on transmittace. 8mm, 11.5mm (used), 15mm, 17mm. |
07/14/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Remeasured effect of off-center placement of crystal on transmittance BUT with removal of all residue/tape off 2x2 crystal face. +1.3°, +2.3°, -1.3°, -2.3. |
07/21/2016 | Summer Students | Light Yield (unmeasured) | Crystal accident , crack on 2x2 face of crystal. |
07/28/2016 | Summer Students | Transverse Transmittance | 200-800nm wavelengths. 10cm down crystal. |
12/20/2016 | TH | Light Yield | run 3507: 11.28 pe/MeV, note: not a good fit to SEP, low stat run. |
07/28/2016 | Summer Students | Longitudinal Transmittance | 200-800nm wavelengths. Sampling period: 0.6s, moving motor at 2mm/s. |
07/07/2017 | Summer Students | Longitudinal Transmittance | 200-800nm wavelengths. Sampling period: 0.6s, moving motor at 2mm/s. |
07/10/2017 | Summer Students | Light Yield | Run 3691. 100ns gw. ~50,000 incidents ~3 layers of teflon tape ~1 layer of electrical. LY=10.01 pe/MeV. |