|07/07/2016 || Summer Students || Longitudinal Transmittance || At 10cm. 200-800nm wavelength. | |07/07/2016 || Summer Students || Longitudinal Transmittance || At 10cm. 200-800nm wavelength. | ||
|- | |- | ||
− | | 07/ | + | | 07/10/2017 || Summer Students || Light Yield || Run 3692. 100ns gw. ~50,000 incidents ~3 layers of teflon tape ~1 layer of electrical. LY=11.00 pe/MeV. |
+ | |- | ||
+ | | 07/26/2017 || Summer Students || Transverse Transmittance Scan || Wavelengths: 200-800 nm. 10cm down.Results discarded due spectrometer error (solved through calibration). | ||
+ | |- | ||
+ | | 07/26/2017 || Summer Students || Xray Radiation || 70 minutes, 160mV, 6mA. ___ mm from beam origin.Results discarded due spectrometer error (solved through calibration). | ||
+ | |- | ||
+ | | 07/26/2017 || Summer Students || Transverse Transmittance Scan || Wavelengths: 200-800 nm. 10cm down.Results discarded due spectrometer error (solved through calibration). | ||
+ | |- | ||
+ | | 07/26/2017 || Summer Students || LED curing || 34 LEDs, 3.5 V for 7 minutes.Results discarded due spectrometer error (solved through calibration). | ||
+ | |- | ||
+ | | 07/26/2017 || Summer Students || Transverse Transmittance Scan || Wavelengths: 200-800 nm. 10cm down.Results discarded due spectrometer error (solved through calibration). | ||
+ | |- | ||
+ | | 07/26/2017 || Summer Students || LED curing || 34 LEDs, 3.5 V for 14 minutes.Results discarded due spectrometer error (solved through calibration). | ||
+ | |- | ||
+ | | 07/26/2017 || Summer Students || Transverse Transmittance Scan || Wavelengths: 200-800 nm. 10cm down.Results discarded due spectrometer error (solved through calibration). | ||
+ | |- | ||
+ | | 07/26/2017 || Summer Students || LED curing || 34 LEDs, 3.5 V for 21 minutes.Results discarded due spectrometer error (solved through calibration). | ||
+ | |- | ||
+ | | 07/26/2017 || Summer Students || Transverse Transmittance Scan || Wavelengths: 200-800 nm. 10cm down. Results discarded due spectrometer error (solved through calibration). | ||
|} | |} | ||
Latest revision as of 10:38, 2 August 2017
⇐ Back to Crystal LOG |
⇐ Back to Neutral Particle Spectrometer |
⇐ Back to the Main_Page |
Time LOG
Date | Experimenter(s) | Description | Comments |
---|---|---|---|
12/14/2015 | Marco | Pictures | Documentation of crystal's conditions as it arrived at CUA. |
01/28/2016 | Marco | Thermal annealing | Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. |
02/25/2016 | Marco | Light Yield | Run 3145: crystal wrapped with ~5 layers of teflon tape and 2 layers of black tape. Steps of 20min, 1cm. T=18oC. LY = 15.97 pe/MeV. |
03/04/2016 | Marco | Transverse Transmittance | At the center of the crystal, every nm from 200nm until 800nm. Sample: 1312 |
03/04/2016 | Marco | Dimensions | Crystal's dimensions measured. Length and six points of width (3 in each face). |
05/23/2016 | Marco | Transverse Transmittance Scan | Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s. |
07/06/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s. |
07/15/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s. |
07/21/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s. |
07/20/2016 | Summer Students | Transverse Transmittance | Wavelengths: 200-800nm. 10cm down crystal. |
07/23/2016 | Summer Students | Light Yield | Run:3242 3 layers of teflon tape and 1 layer of electrical. ~50,000 incidents recorded at ~18 degrees Celsius. |
07/28/2016 | Summer Students | X-ray Irradiation | 160kV, 6.3mA, 70 minutes. 6290R/min ionization rate. |
07/28/2016 | Summer Students | Transverse Transmittance | At 10cm. 200-800nm wavelength. |
07/28/2016 | Summer Students | LED curing | 7 minutes with 34 blue LEDs. 3.5 V. Millipore cover placed over lights and crystal. |
07/28/2016 | Summer Students | Transverse Transmittance | At 10cm. 200-800nm wavelength. |
07/28/2016 | Summer Students | LED curing | 7 more minutes (14 minutes total) with 34 blue LEDs. 3.5 V. Millipore cover placed over lights and crystal. |
07/28/2016 | Summer Students | Transverse Transmittance | At 10cm. 200-800nm wavelength. |
07/28/2016 | Summer Students | LED curing | 7 more minutes (21 minutes total) with 34 blue LEDs. 3.5 V. Millipore cover placed over lights and crystal. |
12/7/2016 | Rich | Thermal annealing | East6, Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. Temperature profile |
12/12/2016 | Rich | Light Yield | Run 3145: crystal wrapped with ~5 layers of teflon tape and 2 layers of black tape. T=18oC. LY ~12 pe/MeV. |
07/07/2016 | Summer Students | Longitudinal Transmittance | At 10cm. 200-800nm wavelength. |
07/10/2017 | Summer Students | Light Yield | Run 3692. 100ns gw. ~50,000 incidents ~3 layers of teflon tape ~1 layer of electrical. LY=11.00 pe/MeV. |
07/26/2017 | Summer Students | Transverse Transmittance Scan | Wavelengths: 200-800 nm. 10cm down.Results discarded due spectrometer error (solved through calibration). |
07/26/2017 | Summer Students | Xray Radiation | 70 minutes, 160mV, 6mA. ___ mm from beam origin.Results discarded due spectrometer error (solved through calibration). |
07/26/2017 | Summer Students | Transverse Transmittance Scan | Wavelengths: 200-800 nm. 10cm down.Results discarded due spectrometer error (solved through calibration). |
07/26/2017 | Summer Students | LED curing | 34 LEDs, 3.5 V for 7 minutes.Results discarded due spectrometer error (solved through calibration). |
07/26/2017 | Summer Students | Transverse Transmittance Scan | Wavelengths: 200-800 nm. 10cm down.Results discarded due spectrometer error (solved through calibration). |
07/26/2017 | Summer Students | LED curing | 34 LEDs, 3.5 V for 14 minutes.Results discarded due spectrometer error (solved through calibration). |
07/26/2017 | Summer Students | Transverse Transmittance Scan | Wavelengths: 200-800 nm. 10cm down.Results discarded due spectrometer error (solved through calibration). |
07/26/2017 | Summer Students | LED curing | 34 LEDs, 3.5 V for 21 minutes.Results discarded due spectrometer error (solved through calibration). |
07/26/2017 | Summer Students | Transverse Transmittance Scan | Wavelengths: 200-800 nm. 10cm down. Results discarded due spectrometer error (solved through calibration). |