HMS AEROGEL DETECTOR (Arthur, Richard, Salim)
 
HMS AEROGEL DETECTOR (Arthur, Richard, Salim)
  
* CUA received 5 Burle 8854 PMTs from former SOS aerogel cherenkov detector (aerogel was n=1.05) - 2 more PMTs and the detector box are at JLab
+
* CUA received 3 Burle [http://www.vsl.cua.edu/cua_phy/images/2/26/Burle_pmt_model.jpeg 8854 PMTs] from former SOS aerogel cherenkov detector (aerogel was n=1.05) - 5 more PMTs and the detector box are at JLab
 
::* These PMTs are about 20 years old
 
::* These PMTs are about 20 years old
::* HV range up to 2500 V, tests at 2200 V show good signals on scope without noise
+
::* HV range up to 2500 V, tests at 2200 V show good signals on scope without noise ([http://www.vsl.cua.edu/cua_phy/images/7/77/Burle_pmts.jpeg Pic1], [http://www.vsl.cua.edu/cua_phy/images/b/b2/Burle_test_setup_jlab.jpeg Pic2], [http://www.vsl.cua.edu/cua_phy/images/b/b8/Burle_signal_scope.jpeg Pic3], [http://www.vsl.cua.edu/cua_phy/images/9/93/Burle_test_setup_jlab_zoom.jpeg Pic4], [http://www.vsl.cua.edu/cua_phy/images/f/fa/Burle_pulse_generator_setting.JPG Pic5], [http://www.vsl.cua.edu/cua_phy/images/5/57/Burle_min_operating_hv.JPG Pic6], [http://www.vsl.cua.edu/cua_phy/images/e/ef/Burle_pmt_signal.jpeg Pic7], [http://www.vsl.cua.edu/cua_phy/images/6/64/Burle_signal_led_and_2mv_background_noise.JPG Pic8])
 
::* Action Item: Need to check single photoelectron and quantum efficiency for these PMTs
 
::* Action Item: Need to check single photoelectron and quantum efficiency for these PMTs
  
* CUA also received 19 XP4500 PMTs, originally from MIT/Bates
+
* CUA also received 19 XP4500 PMTs, originally from MIT/Bates - [http://www.vsl.cua.edu/cua_phy/images/e/e4/List_of_good_PMTs.jpeg List of good PMTs]
 
::* Initial checks in darkbox promising - full tests ongoing
 
::* Initial checks in darkbox promising - full tests ongoing
 
::* PMTs show small after-pulses - can be considered noise. Question: what was the helium source at MIT/Bates?
 
::* PMTs show small after-pulses - can be considered noise. Question: what was the helium source at MIT/Bates?
Line 31: Line 31:
  
  
PION AND KAON STRUCTURE FUNCTIONS (Dannie)
+
[http://www.vsl.cua.edu/cua_phy/images/a/a2/8-2-meetingcopy.pdf PION AND KAON STRUCTURE FUNCTION](Dannie)
  
 
* Completed kinematic studies for kaons, working on pions next
 
* Completed kinematic studies for kaons, working on pions next
Line 40: Line 40:
  
  
IMPACT OF REFLECTOR ON CRYSTAL LIGHT YIELD (Casey)
+
[http://www.vsl.cua.edu/cua_phy/images/3/37/Week_6_Update.pdf IMPACT OF REFLECTOR ON CRYSTAL LIGHT YIELD] (Casey)
  
 
* Tested different materials, difference combinations, and different thicknesses
 
* Tested different materials, difference combinations, and different thicknesses

Latest revision as of 12:19, 3 August 2017

SHMS AEROGEL DETECTOR (Arthur, Andres, Marco,...)

  • Aerogel tray exchange from SP-11 to SP-30 planned for next few days
  • Discussion about events detected from outside the active area for SP-11 aerogel tray
  • Cosmic data suggest that these are scintillation events
  • for comparison, data from full trays show background (not from aerogel) on order 1 photoelectron
  • Action Item: estimate impact of few photoelectron background on particle identification (proton rejection)


HMS AEROGEL DETECTOR (Arthur, Richard, Salim)

  • CUA received 3 Burle 8854 PMTs from former SOS aerogel cherenkov detector (aerogel was n=1.05) - 5 more PMTs and the detector box are at JLab
  • These PMTs are about 20 years old
  • HV range up to 2500 V, tests at 2200 V show good signals on scope without noise (Pic1, Pic2, Pic3, Pic4, Pic5, Pic6, Pic7, Pic8)
  • Action Item: Need to check single photoelectron and quantum efficiency for these PMTs
  • Initial checks in darkbox promising - full tests ongoing
  • PMTs show small after-pulses - can be considered noise. Question: what was the helium source at MIT/Bates?


DIRC (Greg)

  • Discussion about optimizations of the monitoring system for transporting SLAC DIRC bars to JLab
  • prototype built at Indiana
  • expect transport to happen this fall
  • Discussion about sensor tests and lens prototype tests at CERN


PION AND KAON STRUCTURE FUNCTION(Dannie)

  • Completed kinematic studies for kaons, working on pions next
  • Working on offshellness correction
  • Next will start looking at detector simulation to include resolutions


IMPACT OF REFLECTOR ON CRYSTAL LIGHT YIELD (Casey)

  • Tested different materials, difference combinations, and different thicknesses
  • Teflon performs best, mylar not very good
  • Next will test additional materials including Gore reflector (that was used in SHMS aerogel Cherenkov detector construction)
  • Discussion about cutting reflector to shape rather than trying to wrap