Difference between revisions of "MainPage:Nuclear:NPS:PWO:LightTransmittance"
(6 intermediate revisions by the same user not shown) | |||
Line 7: | Line 7: | ||
== Spectrometer benchmark == | == Spectrometer benchmark == | ||
− | A set of tests were performed in order to understand the repetitivity, | + | A set of tests were performed in order to understand the repetitivity, reproducibility, and the effects of some components of the spectrometer in the transmittance measurements of PWO crystals with the Lambda 750 UV/VIS/NIR spectrometer at CUA. |
− | Two types of measurements are planned to be | + | Two types of measurements are planned to be taken with this spectrometer to quantify the crystals transmittance: longitudinal measurement and transverse scan along crystal length. For each of these, a set of benchmark tests were done as described in the following subsections. |
=== Longitudinal Transmittance === | === Longitudinal Transmittance === | ||
Line 38: | Line 38: | ||
<font color="green" size=3>'''→ <u>Position dependence along the compartment</u>'''</font> | <font color="green" size=3>'''→ <u>Position dependence along the compartment</u>'''</font> | ||
− | + | We measured the difference relative to the sample compartment, and along the beam for the PbF2 crystal cube. This test was done to demonstrate | |
Measurements/tests: | Measurements/tests: | ||
Line 67: | Line 67: | ||
<font color="green" size=3>'''→ <u>Interlaboratorial comparison</u>'''</font> | <font color="green" size=3>'''→ <u>Interlaboratorial comparison</u>'''</font> | ||
− | + | Comparing transmittance results with Giessen Univ and others. | |
Measurements/tests: | Measurements/tests: | ||
Line 89: | Line 89: | ||
<font color="green" size=3>'''→ <u>Slit width effects</u>'''</font> | <font color="green" size=3>'''→ <u>Slit width effects</u>'''</font> | ||
− | + | We tested varying slit width effects on transmittance measurements. This was configured via UV WinLab software. | |
Measurements/tests: | Measurements/tests: | ||
Line 98: | Line 98: | ||
|+ | |+ | ||
|- | |- | ||
− | | valign="top"| [[File: | + | | valign="top"| [[File:Slit width config pic.png|thumb|center|340px|A screen shot of the configuration panel for changing the spectrometer slit width in front of the PMT. The nominal slit width is 2.00 nm, as shown on the configuration panel.]] |
|} | |} | ||
Line 111: | Line 111: | ||
<font color="green" size=3>'''→ <u>Compartment windows effects</u>'''</font> | <font color="green" size=3>'''→ <u>Compartment windows effects</u>'''</font> | ||
− | + | We test the effects of changing the compartment windows in the spectrophotometer on our transmittance measurements. | |
Measurements/tests: | Measurements/tests: | ||
Line 137: | Line 137: | ||
<font color="green" size=3>'''→ <u>Final test - interlaboratorial crosscheck</u>'''</font> | <font color="green" size=3>'''→ <u>Final test - interlaboratorial crosscheck</u>'''</font> | ||
− | + | We compare our transmittance results to other universities/labs' measurement of transmittance on the same crystals. | |
Measurements/tests: | Measurements/tests: | ||
Line 159: | Line 159: | ||
=== Transverse Transmittance === | === Transverse Transmittance === | ||
+ | |||
+ | ====Testing the effects of using a black tarp instead of lid to cover compartment to prevent possibilities of escaped light during transverse transmittance measurements==== | ||
<font color="green" size=3>'''→ <u>Cloth to cover the spectrometer</u>'''</font> | <font color="green" size=3>'''→ <u>Cloth to cover the spectrometer</u>'''</font> | ||
− | |||
This test allowed us to make sure there is no effect on the transmittance measurements when working with a cloth covering the sample compartment instead the spectrometer's lid. Using the cloth will be necessary since the 20 cm crystals can only be fully scanned in the spectrometer if the lid is removed, since the sample compartment is not big enough. | This test allowed us to make sure there is no effect on the transmittance measurements when working with a cloth covering the sample compartment instead the spectrometer's lid. Using the cloth will be necessary since the 20 cm crystals can only be fully scanned in the spectrometer if the lid is removed, since the sample compartment is not big enough. | ||
Line 171: | Line 172: | ||
|+ | |+ | ||
|- | |- | ||
− | | valign="top"| [[File:benchm_transverse_BTCM_center.jpg|thumb|center|340px| | + | | valign="top"| [[File:benchm_transverse_BTCM_center.jpg|thumb|center|340px|BTCP crystal positioned at the center of spectrometer.]] |
− | | valign="top"| [[File:Benchm_longitudinal_PbF2_spectrometer.jpg|thumb|center|340px|Spectrometer with | + | | valign="top"| [[File:Benchm_longitudinal_PbF2_spectrometer.jpg|thumb|center|340px|Spectrometer with sample compartment cover.]] |
− | | valign="top"| [[File:benchm_transverse_cloth.jpg|thumb|center|340px|Spectrometer with cloth.]] | + | |} |
+ | {| border="0" style="text-align:center;" width="100%" | ||
+ | |+ | ||
+ | |- | ||
+ | | valign="top"| [[File:benchm_transverse_cloth.jpg|thumb|center|340px|Spectrometer with cloth covering sample compartment (in place of cover).]] | ||
|} | |} | ||
Line 204: | Line 209: | ||
|+ | |+ | ||
|- | |- | ||
− | | valign="top"| [[File:benchm_transverse_plot_slider_noslider.jpg|thumb|center|340px|Measurements of transverse transmittance of | + | | valign="top"| [[File:benchm_transverse_plot_slider_noslider.jpg|thumb|center|340px|Measurements of transverse transmittance of BTCP crystal with and without slider in place.]] |
| valign="top"| [[File:benchm_transverse_diff_slider_noslider.jpg|thumb|center|340px|Difference between these two measurements.]] | | valign="top"| [[File:benchm_transverse_diff_slider_noslider.jpg|thumb|center|340px|Difference between these two measurements.]] | ||
|} | |} | ||
− | <font color="green" size=3>'''→ <u> Actual | + | <font color="green" size=3>'''→ <u> Actual transverse transmittance measurement </u>'''</font> |
− | This test allowed us to evaluate, for the first time, the full setup for transverse transmittance scan at a | + | This test allowed us to evaluate, for the first time, the full setup for transverse transmittance scan at a BTCP crystal. |
Measurements: | Measurements: |
Latest revision as of 15:14, 5 July 2018
⇐ Back to Neutral Particle Spectrometer |
⇐ Back to the Main_Page |
Spectrometer benchmark
A set of tests were performed in order to understand the repetitivity, reproducibility, and the effects of some components of the spectrometer in the transmittance measurements of PWO crystals with the Lambda 750 UV/VIS/NIR spectrometer at CUA.
Two types of measurements are planned to be taken with this spectrometer to quantify the crystals transmittance: longitudinal measurement and transverse scan along crystal length. For each of these, a set of benchmark tests were done as described in the following subsections.
Longitudinal Transmittance
→ Reproducibility
This first test allowed us to check for reproducibility of the measurements in the spectrometer. This was done by placing a crystal at the center of the spectrometer twice, trying to reproduce the same conditions and seeing how the positioning of the crystal in the spectrometer affects our measurements.
Measurements/tests:
- 1 - Measure transmittance with the PbF2 cube at the center of the compartment.
Pictures:
Results:
→ Position dependence along the compartment
We measured the difference relative to the sample compartment, and along the beam for the PbF2 crystal cube. This test was done to demonstrate
Measurements/tests:
- 2 - Measure transmittance with PbF2 cube at 4 additional points along the beam. Compare results to those of the previous point.
Pictures:
Results:
→ Interlaboratorial comparison
Comparing transmittance results with Giessen Univ and others.
Measurements/tests:
- 3 - Measure transmittance of 10-cm CRYTUR crystal installed at the center of the compartment. Compare the result to those from Giessen Univ. and earlier results from summer 2015.
Pictures:
Results:
→ Slit width effects
We tested varying slit width effects on transmittance measurements. This was configured via UV WinLab software.
Measurements/tests:
- 4 - Measure transmittance of 10-cm CRYTUR crystal installed at the center of the compartment with different slit widths (different beam sizes at the center). Compare results to those from previous point.
Pictures:
Results:
→ Compartment windows effects
We test the effects of changing the compartment windows in the spectrophotometer on our transmittance measurements.
Measurements/tests:
- 5 - Measure transmittance of the PbF2 cube and the 10-cm CRYTUR crystal installed at the center of the compartment without the windows. Compare results to those of point 1 and 3.
Pictures:
Results:
→ Final test - interlaboratorial crosscheck
We compare our transmittance results to other universities/labs' measurement of transmittance on the same crystals.
Measurements/tests:
- 6 - Measure transmittance of 20-cm BTCP crystal (windows removed) and compare results with those from Giessen Univ.
Pictures:
File:Benchm longitudinal BTCP.jpg BTCP crystal at spectrometer sample compartment for longitudinal transmittance measurement. |
Results:
Transverse Transmittance
Testing the effects of using a black tarp instead of lid to cover compartment to prevent possibilities of escaped light during transverse transmittance measurements
→ Cloth to cover the spectrometer This test allowed us to make sure there is no effect on the transmittance measurements when working with a cloth covering the sample compartment instead the spectrometer's lid. Using the cloth will be necessary since the 20 cm crystals can only be fully scanned in the spectrometer if the lid is removed, since the sample compartment is not big enough.
Measurements/tests:
- 1 - Measure transmittance of a 20-cm BTCP crystal installed at the center of the compartment with the compartment cover installed and closed.
- 2 - Measure transmittance of 20-cm BTCP crystal installed at the center of the compartment without the compartment cover and a black cloth covering the setup. Do not move the crystal between point (1) and this measurement. This test is done without the diagonal slider. Compare the result with that of the previous point. If there is significant noise in the data, check the positioning of the black cloth.
Pictures:
Results:
→ Slider installation
A slider powered by a stepper motor was configured to position the crystal in the spectrometer compartment for the measurement of transverse transmittance along the crystal. This test allowed us to check for alignment of the crystal to the light beam and for the reproductibility of the measurement using the slider.
Measurements/tests:
- 3 - Install and test the diagonal slider stepper motor with the crystal installed, but no actual transmittance measurement.
- 4 - Measure transmittance of a 20-cm BTCP crystal without the compartment cover, with the black cloth, and with the diagonal slider installed (single crystal position only). If there is significant noise in the data compared to the previous two points, stop to evaluate before proceeding. Check alignment of the crystal to the beam. Check for temporary black cover or paint for the base and any other reflecting surface on the sliding device.
Pictures:
File:Benchm transverse slider1.jpg Slider installed at the spectrometer for transmittance scan along the crystal. |
File:Benchm transverse slider2.jpg Tests of reproductibility with slider in place. |
Results:
File:Benchm transverse plot slider noslider.jpg Measurements of transverse transmittance of BTCP crystal with and without slider in place. |
File:Benchm transverse diff slider noslider.jpg Difference between these two measurements. |
→ Actual transverse transmittance measurement
This test allowed us to evaluate, for the first time, the full setup for transverse transmittance scan at a BTCP crystal.
Measurements:
- 5 - Measure transmittance of a 20-cm BTCP crystal at 5-6 points along the crystal with the diagonal slider (compartment cover off, black cloth installed). Determine the variation of the mean at 50% transmittance.
Pictures:
File:Benchm transverse severalPts1.jpg Slider positioning crystal at edge 1. |
File:Benchm transverse severalPts2.jpg Slider positioning crystal at edge 2. |
Results:
File:Benchm transverse plot severalPts.jpg Transmittance of BTCP crystal at several points along its length, positioned with the slider. |
File:Benchm transverse plot variation.jpg Variation of transmittance... |