Difference between revisions of "MainPage:Nuclear:NPS:PWO:CrystalLOG:SICCAJ19"
Jump to navigation
Jump to search
Line 15: | Line 15: | ||
! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:10%;" | Date | ! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:10%;" | Date | ||
+ | ! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:15%;" | Experimenter(s) | ||
! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:15%;" | Description | ! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:15%;" | Description | ||
! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:75%;" | Comments | ! "background:#f5f5f5;" style="text-align:center; background:lavender; font-weight:bold; width:75%;" | Comments | ||
|- | |- | ||
− | | 12/14/2015 || Pictures || Documentation of crystal's conditions as it arrived at CUA. | + | | 12/14/2015 || Marco || Pictures || Documentation of crystal's conditions as it arrived at CUA. |
|- | |- | ||
− | | 12/15/2015 || Thermal annealing || Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. | + | | 12/15/2015 || Marco || Thermal annealing || Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. |
|- | |- | ||
− | | 12/17/2015 || Transverse Transmittance || Scan at 360, 420, and 620nm. | + | | 12/17/2015 || Marco || Transverse Transmittance || Scan at 360, 420, and 620nm. |
|- | |- | ||
− | | 12/18/2015 || Irradiation || 160 keV photons at 6mA, for 10min. | + | | 12/18/2015 || Marco || Irradiation || 160 keV photons at 6mA, for 10min. |
|- | |- | ||
− | | 12/18/2015 || Transverse Transmittance || Scan at 360, 420, and 620nm after 30 min of irradiation. | + | | 12/18/2015 || Marco || Transverse Transmittance || Scan at 360, 420, and 620nm after 30 min of irradiation. |
|- | |- | ||
− | | 01/06/2016 || LY scan || At 100ns, as described in run 3097. | + | | 01/06/2016 || Marco || LY scan || At 100ns, as described in run 3097. |
|- | |- | ||
− | | 03/04/2016 || Transverse Transmittance || At the center of the crystal, every nm from 200nm until 800nm. Sample: 1314 | + | | 03/04/2016 || Marco || Transverse Transmittance || At the center of the crystal, every nm from 200nm until 800nm. Sample: 1314 |
|- | |- | ||
− | | 03/04/2016 || Dimensions || Crystal's dimensions measured. Length and six points of width (3 in each face). | + | | 03/04/2016 || Marco and Salina || Dimensions || Crystal's dimensions measured. Length and six points of width (3 in each face). |
|- | |- | ||
− | | 05/23/2016 || Transverse Transmittance Scan || Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s. | + | | 05/23/2016 || Marco || Transverse Transmittance Scan || Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s. |
|- | |- | ||
− | | 07/07/2016 || Longitudinal Transmittance Scan || Wavelengths: 200-800 nm. Tested transmittance dependence on which 20x2 face of crystal was face down on the Labjack. | + | | 07/07/2016 || Summer Students || Longitudinal Transmittance Scan || Wavelengths: 200-800 nm. Tested transmittance dependence on which 20x2 face of crystal was face down on the Labjack. |
|- | |- | ||
− | | 07/07/2016 || Longitudinal Transmittance Scan || Wavelengths: 200-800nm. Measured effect of off center crystal placement on transmittance. +1.3°, +2.3°, -1.3°, -2.3°. | + | | 07/07/2016 || Summer Students || Longitudinal Transmittance Scan || Wavelengths: 200-800nm. Measured effect of off center crystal placement on transmittance. +1.3°, +2.3°, -1.3°, -2.3°. |
|- | |- | ||
− | | 07/07/2016 || Longitudinal Transmittance Scan || Wavelengths: 200-800nm. Measured effect of iris diameter on transmittance. 8.0mm, 11.5mm(used), 15mm, and 17mm. | + | | 07/07/2016 || Summer Students || Longitudinal Transmittance Scan || Wavelengths: 200-800nm. Measured effect of iris diameter on transmittance. 8.0mm, 11.5mm(used), 15mm, and 17mm. |
|- | |- | ||
− | | 07/ | + | | 07/07/2016 || Summer Students || Longitudinal Transmittance Scan || Wavelengths: 200-800nm. Remeasured effect of off center crystal placement on transmittance BUT with removal of all tape/residue of 2x2 crystal face. +1.3°, +2.3°, -1.3°, -2.3°. |
|- | |- | ||
− | | 07/ | + | | 07/11/2016 || Summer Students || Longitudinal Transmittance Scan || Wavelengths: 200-800 nm. Measured effect of which face of crystal faced light beam in spectrometer. "flip" of crystal. |
|- | |- | ||
− | | 07/26/2016 || Light Yield || Run:3248 3 layers of teflon tape and 1 layer of elecrtical. ~50,000 incidents recorded at ~18 degrees Celsius. | + | | 07/26/2016 || Summer Students || Light Yield || Run:3248 3 layers of teflon tape and 1 layer of elecrtical. ~50,000 incidents recorded at ~18 degrees Celsius. |
|- | |- | ||
− | | 07/26/2016 || Transverse Transmittance || 200nm-800nm. 10cm down crystal. | + | | 07/26/2016 || Summer Students || Transverse Transmittance || 200nm-800nm. 10cm down crystal. |
|- | |- | ||
− | | 11/29/2016 || Light Yield || Run:3421 3 layers of teflon tape and 1 layer of elecrtical. ~60,000 incidents recorded at ~18 degrees Celsius. | + | | 11/29/2016 || Richard || Light Yield || Run:3421 3 layers of teflon tape and 1 layer of elecrtical. ~60,000 incidents recorded at ~18 degrees Celsius. |
|} | |} | ||
Revision as of 20:25, 30 November 2016
⇐ Back to Crystal LOG |
⇐ Back to Neutral Particle Spectrometer |
⇐ Back to the Main_Page |
Time LOG
Date | Experimenter(s) | Description | Comments |
---|---|---|---|
12/14/2015 | Marco | Pictures | Documentation of crystal's conditions as it arrived at CUA. |
12/15/2015 | Marco | Thermal annealing | Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. |
12/17/2015 | Marco | Transverse Transmittance | Scan at 360, 420, and 620nm. |
12/18/2015 | Marco | Irradiation | 160 keV photons at 6mA, for 10min. |
12/18/2015 | Marco | Transverse Transmittance | Scan at 360, 420, and 620nm after 30 min of irradiation. |
01/06/2016 | Marco | LY scan | At 100ns, as described in run 3097. |
03/04/2016 | Marco | Transverse Transmittance | At the center of the crystal, every nm from 200nm until 800nm. Sample: 1314 |
03/04/2016 | Marco and Salina | Dimensions | Crystal's dimensions measured. Length and six points of width (3 in each face). |
05/23/2016 | Marco | Transverse Transmittance Scan | Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s. |
07/07/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Tested transmittance dependence on which 20x2 face of crystal was face down on the Labjack. |
07/07/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800nm. Measured effect of off center crystal placement on transmittance. +1.3°, +2.3°, -1.3°, -2.3°. |
07/07/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800nm. Measured effect of iris diameter on transmittance. 8.0mm, 11.5mm(used), 15mm, and 17mm. |
07/07/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800nm. Remeasured effect of off center crystal placement on transmittance BUT with removal of all tape/residue of 2x2 crystal face. +1.3°, +2.3°, -1.3°, -2.3°. |
07/11/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Measured effect of which face of crystal faced light beam in spectrometer. "flip" of crystal. |
07/26/2016 | Summer Students | Light Yield | Run:3248 3 layers of teflon tape and 1 layer of elecrtical. ~50,000 incidents recorded at ~18 degrees Celsius. |
07/26/2016 | Summer Students | Transverse Transmittance | 200nm-800nm. 10cm down crystal. |
11/29/2016 | Richard | Light Yield | Run:3421 3 layers of teflon tape and 1 layer of elecrtical. ~60,000 incidents recorded at ~18 degrees Celsius. |