+ | | 12/20/2016 || TH || Light Yield || already wrapped crystal, run 3502 start at 20degC and with source at same location as for 20-cm long crystals, result: 15.4 pe/MeV, run 3503 (better thermally equlibrated): 15.66 pe/MeV, run 3504 is same as before, but with other end of crystal taped shut, result: 14.76 pe/MeV, run 3505 is same as 3504, but with the source aligned to crystal center using stepper motor, result: 15.7 pe/MeV. | ||
+ | |||
|} | |} | ||
Revision as of 19:04, 8 January 2017
⇐ Back to Crystal LOG |
⇐ Back to Neutral Particle Spectrometer |
⇐ Back to the Main_Page |
Time LOG
Date | Experimenter(s) | Description | Comments |
---|---|---|---|
xx/xx/2015 | Marco | Wrapped | Preparing for light yield measurement. Approximately 5 layers of teflon tape, and 2 layers of black tape. |
xx/xx/2015 | Marco | LY scan | ... |
12/14/2015 | Marco | unwrapped | Preparing for annealing. |
12/14/2015 | Marco | Thermal annealing | Starting 11:00AM. Ramp up at 18oC/hour until 200oC, stay at 200oC for 10 hours, ramp down at -18oC/hour. |
12/17/2015 | Marco | Longitudinal Transmittance | After annealing. |
07/15/2016 | Summer Students | Longitudinal Transmittance | Wavelength:200-800nm. |
08/11/2016 | Summer Students | LY scan | RUN:3411. Motor was lowered to approximately 5 cm down crystal. ~18oC. LY: 11.986 phe/MeV- |
11/18/2016 | Rich | Light Yield Scan | Run 3422: ~100 events; 10 cm crytur crystal wrapped with ~4 layers of teflon tape and 2 layers of black electrical tape. T=18°C. LY = 9.3978 pe/Mev (see pictures below). Not enough runs? Note: the error bars used were not the 0.5 standard used previously, but instead the calculated uncertainty for each run. |
11/29/2016 | Salina | Light Yield Scan | Run 3432: ~100K events; 10 cm crytur crystal wrapped with ~5 layers of teflon tape and 2 layers of black electrical tape. T=18°C. LY = 8.23 ± 1.08 pe/MeV. Not enough statistics? Need to re-run with more stats (see pictures below) |
11/30/2016 | Salina | Light Yield Scan | Run 3434: ~1.4M events; Re-run of 10 cm crytur crystal wrapped with ~5 layers of teflon tape and 2 layers of black electrical tape. T=18°C. LY = 9.67 ± 0.53 pe/MeV (see pictures below). |
12/02/2016 | Rich | Longitudinal Transmittance | Measured center of crystal before annealing. |
12/20/2016 | TH | Light Yield | already wrapped crystal, run 3502 start at 20degC and with source at same location as for 20-cm long crystals, result: 15.4 pe/MeV, run 3503 (better thermally equlibrated): 15.66 pe/MeV, run 3504 is same as before, but with other end of crystal taped shut, result: 14.76 pe/MeV, run 3505 is same as 3504, but with the source aligned to crystal center using stepper motor, result: 15.7 pe/MeV. |