Difference between revisions of "MainPage:Nuclear:NPS:PWO:CrystalLOG:SICCAJ20"
Jump to navigation
Jump to search
Line 87: | Line 87: | ||
| valign="top"| | | valign="top"| | ||
[[File:J20-100nsgw.png|thumb|left|340px| 07/23/2016]] | [[File:J20-100nsgw.png|thumb|left|340px| 07/23/2016]] | ||
+ | |} | ||
+ | |||
+ | == Irradiation Studies == | ||
+ | {|border="0" style="text-align:center;" width="100%" | ||
+ | |- | ||
+ | | valign="top"| | ||
+ | [[File:J20IR.png|thumb|left|340px| 07/27/2017]] | ||
+ | |} | ||
+ | |||
+ | == LED Curing == | ||
+ | {|border="0" style="text-align:center;" width="100%" | ||
+ | |- | ||
+ | | valign="top"| | ||
+ | [[File:J20LED.png|thumb|left|340px| 07/27/2017]] | ||
|} | |} |
Latest revision as of 10:30, 2 August 2017
⇐ Back to Crystal LOG |
⇐ Back to Neutral Particle Spectrometer |
⇐ Back to the Main_Page |
Time LOG
Date | Experimenter(s) | Description | Comments |
---|---|---|---|
12/14/2015 | Marco | Pictures | Documentation of crystal's conditions as it arrived at CUA. |
12/15/2015 | Marco | Thermal annealing | Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. |
12/17/2015 | Marco | Transverse Transmittance | Scan at 360, 420, and 620nm. |
12/18/2015 | Marco | Irradiation | 160 keV photons at 6mA, for 10min. |
12/18/2015 | Marco | Transverse Transmittance | Scan at 360, 420, and 620nm after 30 min of irradiation. |
01/07/2016 | Marco | LY scan | LY measurement with window width of 100ns, at 18oC (see run 3098). LY = 14.76 pe/MeV. |
03/04/2016 | Marco | Transverse Transmittance | At the center of the crystal, every nm from 200nm until 800nm. Sample: 1313 |
03/04/2016 | Marco and Salina | Dimensions | Crystal's dimensions measured. Length and six points of width (3 in each face). |
05/23/2016 | Marco | Transverse Transmittance Scan | Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s. |
07/08/2016 | Marco | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Measured effect of off-center placement of crystal on transmittance. +1.3°, +2.3°, -1.3°, -2.3. |
07/08/2016 | Summer Students | Longitudinal Transmittance Scan | Wavelengths: 200-800 nm. Measured light beam entering both end face of crystal in spectrometer. "flip" of crystal. |
07/23/2016 | Summer Students | Light Yield | Run: 3246. 3 layers of teflon tape and 1 of electrical. ~50.000 incidents and ~18 degrees Celsius |
07/28/2016 | Summer Students | Transverse Transmittance | At 10cm. 200-800nm wavelength. |
07/28/2016 | Summer Students | X-Ray Irradiation | 160kV 6.3mA 70 minutes. 14.16 cm for LabJack on shelf 8. 6290R/min ionization rate. |
07/28/2016 | Summer Students | Transverse Transmittance (spectrometer failure) | Unsuccessful measurement of transmittance. |
07/07/2017 | Summer Students | Transverse Transmittance | 200-800nm wavelengths. Sampling period: 0.6s, moving motor at 2mm/s. |
07/07/2017 | Summer Students | Irradiation | 160 keV photons at 6.3mA, for 90min. (Unsuccessful due to fuse shorting) |
07/27/2017 | Summer Students | Transverse Transmittance Scan | Wavelengths: 200-800 nm. 10cm down. |
07/27/2017 | Summer Students | Xray Radiation | 70 minutes, 160mV, 6mA. ___ mm from beam origin. |
07/27/2017 | Summer Students | Transverse Transmittance Scan | Wavelengths: 200-800 nm. 10cm down. |
07/27/2017 | Summer Students | LED curing | 34 LEDs, 3.5 V for 7 minutes. |
07/27/2017 | Summer Students | Transverse Transmittance Scan | Wavelengths: 200-800 nm. 10cm down. |
07/27/2017 | Summer Students | LED curing | 34 LEDs, 3.5 V for 7 minutes. |
07/27/2017 | Summer Students | Transverse Transmittance Scan | Wavelengths: 200-800 nm. 10cm down. |
07/27/2017 | Summer Students | LED curing | 34 LEDs, 3.5 V for 7 minutes. |
07/27/2017 | Summer Students | Transverse Transmittance Scan | Wavelengths: 200-800 nm. 10cm down. |