MainPage:Nuclear:NPS:PWO:LightTransmittance
Jump to navigation
Jump to search
⇐ Back to Neutral Particle Spectrometer |
⇐ Back to the Main_Page |
Spectrometer benchmark
A set of tests were performed in order to understand the repetitivity, reprodutibility, and the effects of some components of the spectrometer in the transmittance measurements of PWO crystals with the Lambda 750 UV/VIS/NIR spectrometer at CUA.
Two types of measurements are planned to be performed with this spectrometers to quantify the crystals transmittance: longitudinal measurement and transverse scan along crystal length. For each of these, a set of benchmark tests were done as described in the following subsections.
Transverse Transmittance
- Measure transmittance of a 20-cm BTCP crystal installed at the center of the compartment with the compartment cover installed and closed