Time LOG

CRYTUR 01
Date Experimenter(s) Description Comments
xx/xx/2015 Marco Wrapped Preparing for light yield measurement. Approximately 5 layers of teflon tape, and 2 layers of black tape.
xx/xx/2015 Marco LY scan ...
12/14/2015 Marco unwrapped Preparing for annealing.
12/14/2015 Marco Thermal annealing Starting 11:00AM. Ramp up at 18oC/hour until 200oC, stay at 200oC for 10 hours, ramp down at -18oC/hour.
12/17/2015 Marco Longitudinal Transmittance After annealing.
07/15/2016 Summer Students Longitudinal Transmittance Wavelength:200-800nm.
08/11/2016 Summer Students LY scan RUN:3411. Motor was lowered to approximately 5 cm down crystal. ~18oC. LY: 11.986 phe/MeV-
11/18/2016 Rich Light Yield Scan Run 3422: ~100 events; 10 cm crytur crystal wrapped with ~4 layers of teflon tape and 2 layers of black electrical tape. T=18°C. LY = 9.3978 pe/Mev (see pictures below). Not enough runs? Note: the error bars used were not the 0.5 standard used previously, but instead the calculated uncertainty for each run.
11/29/2016 Salina Light Yield Scan Run 3432: ~100K events; 10 cm crytur crystal wrapped with ~5 layers of teflon tape and 2 layers of black electrical tape. T=18°C. LY = 8.23 ± 1.08 pe/MeV. Not enough statistics? Need to re-run with more stats (see pictures below)
11/30/2016 Salina Light Yield Scan Run 3434: ~1.4M events; Re-run of 10 cm crytur crystal wrapped with ~5 layers of teflon tape and 2 layers of black electrical tape. T=18°C. LY = 9.67 ± 0.53 pe/MeV (see pictures below).
12/02/2016 Rich Longitudinal Transmittance Measured center of crystal before annealing.

Transmittance measurements

Longitudinal transmittance measured at Lambda 750 (no integrating sphere).
Longitudinal transmittance measured at Lambda 750 (no integrating sphere).

Pictures at 12/14/2015

CRYTUR shown inside oven for thermal annealing.

Light Yield

Run 3422 taken on 11/18/2016: Fit of histogram with ~100K events.
Run 3422 taken on 11/18/2016: LY as a function of position. Error bars based off calculatd uncertainty.
Run 3432 taken on 11/29/2016: Fit of histogram with ~100K events.
Run 3432 taken on 11/29/2016: LY as a function of position.
Run 3434 taken on 11/29/2016: Fit of histogram with ~1.4M events.
Run 3434 taken on 11/29/2016: LY as a function of position.

Additional Data/Info

Graph showing CRYTUR crystal longitudinal measurements before annealing (red) and after annealing (green).
Graph showing the difference of the CRYTUR crystal longitudinal measurements before and after annealing.