Date Description Comments
12/14/2015 Pictures Documentation of crystal's conditions as it arrived at CUA.
12/15/2015 Thermal annealing Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour.
12/17/2015 Transverse Transmittance Scan at 360, 420, and 620nm.
12/18/2015 Irradiation 160 keV photons at 6mA, for 10min.
12/18/2015 Transverse Transmittance Scan at 360, 420, and 620nm after 30 min of irradiation.
01/07/2016 LY scan LY measurement with window width of 100ns, at 18oC (see run 3098). LY = 14.76 pe/MeV.
03/04/2016 Transverse Transmittance At the center of the crystal, every nm from 200nm until 800nm. Sample: 1313
03/04/2016 Dimensions Crystal's dimensions measured. Length and six points of width (3 in each face).
05/23/2016 Transverse Transmittance Scan Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s.
07/05/2016 Longitudinal Transmittance Scan Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s.