Time LOG

SICCA J16
Date Description Comments
12/14/2015 Pictures Documentation of crystal's conditions as it arrived at CUA.
12/14/2015 Wrapped Preparing for light yield measurement. Approximately 5 layers of teflon tape, and 2 layers of black tape.
12/15/2015 LY scan Measurement of LY along crystal. More details in the LOGBOOK, runs xx..
02/01/2016 Thermal annealing Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour.
03/04/2016 Transverse Transmittance At the center of the crystal, every nm from 200nm until 800nm. Sample: 1317
03/04/2016 Dimensions Crystal's dimensions measured. Length and six points of width (3 in each face).
05/23/2016 Transverse Transmittance Scan Wavelengths: 360nm, 420nm, 620nm. Sampling period: 0.6s, moving motor at 2mm/s.
07/05/2016 Longitudinal Transmittance Scan Wavelengths: 200-800nm. Measured effect of iris diameter on transmittance. 8.0mm, 11.5mm, 15mm, and 17mm.
07/14/2016 Longitudinal Transmittance Scan Wavelengths: 200-800nm. Measured effect of off center crystal placement on transmittance. +1.3°C, +2.3°C, -1.3°C, -2.3°C.
07/26/2016 Light Yield Run 3250. 100ns gw. ~50,000 incidents ~3 layers of teflon tape ~1 layer of electrical.
07/29/2016 Transverse Transmittance At 10cm. 200-800nm wavelength.
07/29/16 X-Ray Irradiation 160kV, 6.3mA, 70 minutes. 6290 R/min.
07/29/2016 Transverse Transmittance At 10cm. 200-800nm wavelength.
07/29/2016 LED curing 7 minutes with 34 blue LEDs. 3.5 V. Millipore cover placed over lights and crystal.
07/29/2016 Transverse Transmittance At 10cm. 200-800nm wavelength.
07/29/2016 LED curing 7 more minutes (14 minutes total) with 34 blue LEDs. 3.5 V. Millipore cover placed over lights and crystal.
07/29/2016 Transverse Transmittance At 10cm. 200-800nm wavelength.
07/29/2016 LED curing 7 more minutes (21 minutes total) with 34 blue LEDs. 3.5 V. Millipore cover placed over lights and crystal.

Pictures at 12/14/2015

xxxx.
xxxx.
xxxx.
xxxx.

Transmittance Uncertainty 07/2016

Iris diameter vs Transmittance.
Ideal crystal placement would be 20cm side parallel to light beam. Reasonable placement errors would be 1.3° and 2.3° off of this. At these angles the beam still appeared to fall within the crystla face under the "Align" function

Light Yield

Iris diameter vs Transmittance.

Radiation Hardness

Iris diameter vs Transmittance.

LED curing

Iris diameter vs Transmittance.


|}