MainPage:Nuclear:NPS:PWO:CrystalLOG:SICCAJ21

From cua_phy
Jump to navigation Jump to search
⇐ Back to Crystal LOG
⇐ Back to Neutral Particle Spectrometer
⇐ Back to the Main_Page

Time LOG

SICCA J21
Date Description Comments
12/14/2015 Pictures Documentation of crystal's conditions as it arrived at CUA.
01/28/2016 Thermal annealing Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour.
02/25/2016 Light Yield Run 3145: crystal wrapped with ~5 layers of teflon tape and 2 layers of black tape. Steps of 20min, 1cm. T=18oC. LY = 15.97 pe/MeV.
03/04/2016 Transverse Transmittance At the center of the crystal, every nm from 200nm until 800nm. Sample: 1312
03/04/2016 Dimensions Crystal's dimensions measured. Length and six points of width (3 in each face).
05/23/2016 Transverse Transmittance Scan Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s.
07/06/2016 Longitudinal Transmittance Scan Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s.
07/15/2016 Longitudinal Transmittance Scan Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s.
07/21/2016 Longitudinal Transmittance Scan Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s.
07/21/2016 Light Yield Run:3236 3 layers of teflon tape and 1 layer of elecrtical. ~50,000 incidents recorded at ~18 degrees Celsius.

Pictures

12/14/2015.
12/14/2016.
Error creating thumbnail: Unable to save thumbnail to destination
|| 7/18/16 The lid of Spectrometer fell on crystal. Dimensions of chip are 15.88x10.15x9.62mm