12/09/2015 | Marco | unwrapped | Unwrapped as arrived at CUA. Cracks observed on the face - see Pictures 12/09/2015. |
12/10/2015 | Marco | Thermal annealing | Starting 10:30AM. Ramp up at 18oC/hour until 200oC, stay at 200oC for 10 hours, ramp down at -18oC/hour. |
12/11/2015 | Marco | Wrapped | Preparing for light yield measurement. Approximately 5 layers of teflon tape, and 2 layers of black tape - see Pictures 12/11/2015. |
12/14/2015 | Marco | LY scan | Measurement of LY along crystal. Cracked area far apart from PMT; cracked face opposite to the radioactive side. More details in the LOGBOOK, runs 2942.. |
12/xx/2015 | Marco | Thermal annealing | Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. |
12/17/2015 | Marco | Long. Transmittance | Crystal wrapped. |
12/17/2015 | Marco | Unwrapped | Preparing for transverse transmittance. |
12/17/2015 | Marco | Transverse Transmittance | Scan at 360, 420, and 620nm. |
12/18/2015 | Marco | Irradiation | 160 keV photons at 6mA, for 10min. |
12/18/2015 | Marco | Transverse Transmittance | Scan at 360, 420, and 620nm after 30 min from irradiation. |
01/14/2016 | Marco | Transverse Transmittance | Scan at 420nm (scan setup realigned). |
01/14/2016 | Marco | Irradiation | 160 keV photons at 6mA, for 20min. |
01/14/2016 | Marco | Transverse Transmittance | Scan at 420nm after 30 min from irradiation. |
01/14/2016 | Marco | Transverse Transmittance | Spectral measurement. |
02/01/2016 | Marco | Thermal annealing | Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. |
05/16/2016 | Salina | Transverse Transmittance | Scan at 420nm, 360nm, 650nm. |
05/16/2016 | Marco | Irradiation | 700 mGy of Co-60 at dose rates of ~70 mGy/h (10 h) - source at 3cm from crystal. Cylinder of 1 inch per 2cm irradiated. |
05/17/2016 | Salina | Transverse Transmittance | Scan at 420nm after irradiation. |
05/17/2016 | Marco | Irradiation | 840 mGy of Co-60 at dose rates of ~70 mGy/h (12 h). Cylinder of 1 inch per 2cm irradiated. |
05/18/2016 | Marco | Irradiation | 11200 mGy (total 12Gy with previous irrad) of Co-60 at dose rates of ~105 mGy/h (xx h) - source at 1cm from crystal. Cylinder of 1 inch per 2cm irradiated. |
05/24/2016 | Marco | Transverse Transmittance | Spectral, at center (position where irradiated with Co-60). In preparation for x-ray machine. |
05/24/2016 | Marco | Irradiation (x-ray) | 90min @ 160kV @ 6mA. |
05/24/2016 | Marco | Transverse Transmittance | At same position as before, after about 5min of irradiation in x-ray machine. |
05/24/2016 | Marco | Irradiation (x-ray) | 180min @ 160kV @ 6mA. |
06/03/2016 | Marco | Thermal annealing | Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour. |
06/06/2016 | Marco | Transverse Transmittance | Spectral measurement with Lambda 750. |
06/06/2016 | Marco | Transverse Transmittance | Spectral measurement with Lambda 950 - integrated sphere (may not be at exactly the same position). |
06/06/2016 | Marco | Crystal wrapped | 5 layers of teflon tape and 2 layers of black electrical tape. |
06/06/2016 | Marco | Longitudinal Transmittance | Spectral measurement with Lambda 950 - integrated sphere. |