DETECTORS AT JLAB
SHMS AEROGEL CHERENKOV
- Discussion about recent cosmic run results and PMT response
- Noise in raw time spectra - possibly due to noise that was not visible during KPP because of smaller time window (160 ns vs 400 ns) and cosmic vs. real data taking differences (e.g. tracking requirements)
- One PMT (channel #6 negative - see slides) needs to be replaced. Use one of the XP4572 from detector that was at CERN (List of these PMTs).
NEUTRAL PARTICLE SPECTROMETER
- Purchase order for dimension measurement instrument submitted
- Discussion that need to measure dimensions of both, crystals and PMTs (requirement on PMTs is 19.2mm diameter)
- Work on database to store individual crystal information ongoing
- Discussion of crystal characterization results from about 20% of the crystals
DIRC
- Making progress - in process of preparing bars for transportation
CERN DETECTORS
- Discussion about ongoing Fermilab based statistics course and grant writing activities
- Discussion about instrumentation for CUA clean room
DATA ANALYSIS
- Preparing for APS DNP meeting