MainPage:Nuclear:KaonDetector:PMTCharacteristics:LogBook
Jump to navigation
Jump to search
⇐ Back to the Main_Page |
⇐ Neutral Particle Spectrometer |
LogBook on the CODA data acquisition, at CUA
Page (last), [6], [5], [4], [3], [2], [1].
Run Number | Date | User | PMT | Crystal | High Voltage | Gate (ns) | Experiment | Description and observations | Image | |
---|---|---|---|---|---|---|---|---|---|---|
2759 | 11/12/2015 | Marco | XP2262 (SN 34426) | - | -2.300 (NIM) | 100 | Gain | LED (green) gain calibration for NPS light yield. Na22 source is also inside, so may have noise. Rate of ~6.5kHz. T=18oC. SEP: 66.67 (fit range 50-85). PED: 27.66 (fit range 25-35). | Error creating thumbnail: Unable to save thumbnail to destination
| |
2760 | 11/12/2015 | Marco | XP2262 (SN 34426) | - | -2.300 (NIM) | 100 | Gain | Same as 2759, reduce light intensity. SEP: 63.92 (fit range 50-85). PED: 27.65 (fit range 25-35). | Error creating thumbnail: Unable to save thumbnail to destination
| |
2761 | 11/12/2015 | Marco | XP2262 (SN 34426) | - | -2.300 (NIM) | 100 | Gain | Same as 2760; source removed from dark box. SEP: 67.41 (fit range 50-85). PED: 27.63 (fit range 25-35). | Error creating thumbnail: Unable to save thumbnail to destination
| |
2762 | 11/12/2015 | Marco | XP2262 (SN 34426) | - | -2.300 (NIM) | 100 | Gain | Same as 2761; slightly higher light intensity. SEP: 67.78 (fit range 50-85). PED: 27.72 (fit range 25-35). | Error creating thumbnail: Unable to save thumbnail to destination
| |
2763 | 11/12/2015 | Marco | XP2262 (SN 34426) | - | -2.300 (NIM) | 100 | Gain | Light intensity brought to zero (width=0). SEP: 61.98 (fit range 50-85). PED: 27.63 (fit range 25-35). | Error creating thumbnail: Unable to save thumbnail to destination
| |
2766 | 11/12/2015 | Marco | XP2262 (SN 34426) | BTCP ### | -2.300 (NIM) | 100 | Light Yield | Crystal BTCP ###, wrapped with specular reflector (low statistics - 1st test). Na-22 source. T=~13oC (thermometer not in the right position). Mean peak: 353.40 (fit range 170-800). LY=18.0 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
2767 | 11/12/2015 | Marco | XP2262 (SN 34426) | BTCP ### | -2.300 (NIM) | 100 | Light Yield | Same as 2766, high statistics (~900k evts). Peak center: 329.02 (fit range 170-800). LY_fit= pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
2768 | 11/17/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 100 | Light Yield | CRYTUR 10cm sample crystal, wrapped with specular reflector. 18 oC (thermometer positioned). Peak center: 286.78 (fit range 170-400). SEP: 70.35 (fit range 50-85). PED: 27.36 (fit range 25-35). LY_fit=11.8 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
2769 | 11/18/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 100 | Light Yield | CRYTUR 10cm sample crystal, wrapped with ~5 layers of teflon. T=18oC (2 hours of stabilization only). Peak center: 444.07 (fit range 170-800). SEP and PED: same as 2768 (not enough stats.) LY_fit= 18.97 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
2771 | 11/19/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 100 | LY | Same as 2769, with temperature stable for the whole night; 900k evts. | Error creating thumbnail: Unable to save thumbnail to destination
| |
2774..2779 | 11/24/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 100 | LY scan | CRYTUR 10 cm scan. Same configuration as 2769. Running 10 minutes per position, in steps of 1cm. | File:2774.png | |
2781..2793 | 11/30/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 100 | LY scan | CRYTUR 10 cm scan, temperature NOT stable (just testing CODA and scan auto mode). Same configuration as 2769. Running 10 minutes per position, in steps of 1cm. | File:2781.png | |
2797..2800 | 12/02/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 100 | LY scan | CRYTUR 10 cm scan, T=18.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm. | File:2797.png | |
2803..2815 | 12/02/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 100 | LY scan | Repeated 2797 scan without CODA crash! CRYTUR 10 cm scan, T=18.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm. LY=19.77 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
2821..2832 | 12/03/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 100 | LY scan | CRYTUR 10 cm scan, T=10.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm, 12 positions. LY=22.99 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
2834 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 36.3 | LY timing | CRYTUR 10 cm, T=10.0 oC. Position center (5cm). Using gain from combination of runs 2821..2832: SEP-PED=. LY is a result of a gaussian fit in the range 170-1000. Gate left edge at 48ns before first signal in oscilloscope histogram. (gate delay adjusted to have minimum with no signal inside) | File:2834.png | |
2835 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 46.5 | LY timing | Same as 2834, adjusted gate width only. | File:2835.png | |
2836 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 56.1 | LY timing | Same as 2834, adjusted gate width only. | File:2836.png | |
2837 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 66.7 | LY timing | Same as 2834, adjusted gate width only. | File:2837.png | |
2838 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 76.1 | LY timing | Same as 2834, adjusted gate width only. | File:2838.png | |
2839 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 86.6 | LY timing | Same as 2834, adjusted gate width only. | File:2839.png | |
2840 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 96.7 | LY timing | Same as 2834, adjusted gate width only. | File:2840.png | |
2841 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 106.6 | LY timing | Same as 2834, adjusted gate width only. | File:2841.png | |
2842 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 117.0 | LY timing | Same as 2834, adjusted gate width only. | File:2842.png | |
2843 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 125.9 | LY timing | Same as 2834, adjusted gate width only. | File:2843.png | |
2844 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 1004 | LY timing | Same as 2834, adjusted gate width only. | File:284.png | |
2834..2844 | 12/04/2015 | Marco | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | -- | LY timing | Plot of runs 2834..2844. | File:2834 2844.png | |
2846..2868 | 12/07/2015 | Marco | XP2262 (SN 34426) | BTCP ### | -2.300 (NIM) | 149 (eff. 100) | LY scan | BTCP 20cm ###, wrapped with ~5 layers of teflon and black tape. T=17oC (NOT stable, test only). Gate width is 149ns, but ~49ns is before the begining of signal, so effective gate width is 100ns. Steps of 1cm, 5min in each position. PROBLEM in some runs... to many counts? BAD CONNECTOR! | File:2846.png | |
2890..2911 | 12/07/2015 | Marco | XP2262 (SN 34426) | BTCP ### | -2.300 (NIM) | 149 (eff. 100) | LY scan | Same as 2846, but T=18oC (stable) and 105min in each position - total ~40h scan. LY = 12.85 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
2914 | 12/14/2015 | Marco | XP2262 (SN 34426) | SICCA 02 | -2.300 (NIM) | 149 (eff. 100) | LY | SICCA 02 crystal wrapped with ~5 layers of teflon tape and 2 layers of black tape. Cracked face placed opposite to the radioactive source, and to the top (far apart from PMT). Event rate of approx. 67Hz. Temperature almost stable at 18oC, run simple test before starting scan. | Error creating thumbnail: Unable to save thumbnail to destination
| |
2915..2930 | 12/14/2015 | Marco | XP2262 (SN 34426) | SICCA 02 | -2.300 (NIM) | 149 (eff. 100) | LY scan | Same as 2914. Temperature almost stable at 18oC, scan of 5min per step, every 1cm, to test before long scan (interrupted on purpose to start long run). | File:2915.png | |
2932..2939 | 12/14/2015 | Marco | XP2262 (SN 34426) | SICCA 02 | -2.300 (NIM) | 149 (eff. 100) | LY scan | Same as 2914. T=18oC, scan of 50min per step, every 1cm. CODA crashed. LY=16.83 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
2941..xx | 12/15/2015 | Marco | XP2262 (SN 34426) | SICCA 02 | -2.300 (NIM) | 149 (eff. 100) | LY scan | Same as 2932, but steps of 10min. T=18oC. CODA crashed. LY=15.94 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
2961..2963 | 12/15/2015 | Marco | XP2262 (SN 34426) | SICCA J16 | -2.300 (NIM) | 149 (eff. 100) | LY scan | SICCA J16 crystal wrapped with ~5 layers of teflon tape and 2 layers of black tape. Steps of 105min, 1cm. T=18oC. CODA crashed. | File:2961.png | |
2966..2978 | 12/17/2015 | Marco | XP2262 (SN 34426) | SICCA J16 | -2.300 (NIM) | 149 (eff. 100) | LY scan | Same as 2961, but with steps of 10min. CODA crashed. LY=16.46 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
2981..2987 | 12/17/2015 | Marco | XP2262 (SN 34426) | SICCA J16 | -2.300 (NIM) | 149 (eff. 100) | LY scan | Same as 2961, but with steps of 50min. CODA crashed. LY=17.22 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3030..3044 | 01/04/2016 | Marco | XP2262 (SN 34426) | SICCA J18 | -2.300 (NIM) | 149 (eff. 100) | LY scan | SICCA J18 crystal wrapped with ~5 layers of teflon tape and 2 layers of black tape. Steps of 50min, 1cm. T=18oC. CODA crashed. LY=13.54 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3047..3066 | 01/05/2016 | Marco | XP2262 (SN 34426) | SICCA J18 | -2.300 (NIM) | 149 (eff. 100) | LY scan | Same as 3029, but with steps of 10min. LY=13.75 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3070..30xx | 01/05/2016 | Marco | XP2262 (SN 34426) | SICCA J19 | -2.300 (NIM) | 149 (eff. 100) | LY scan | SICCA J19 crystal wrapped with ~5 layers of teflon tape and 2 layers of black tape (new tape in use, 2in wide and not so straight). Steps of 50min, 1cm. T=18oC. CODA crashed. | File:3070.png | |
3097 | 01/06/2016 | Marco | XP2262 (SN 34426) | SICCA J19 | -2.300 (NIM) | 149 (eff. 100) | LY scan | Same as 2070. As CODA keeps crashing, decided to take one run only of 220 min for the 22 positions, 10min in each position. Analysis done taking the mean rate and considering the number of events in 10min interval from that rate for each step. LY= 19.13 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3098 | 01/07/2016 | Marco | XP2262 (SN 34426) | SICCA J20 | -2.300 (NIM) | 149 (eff. 100) | LY scan | SICCA J20 crystal wrapped with ~5 layers of teflon tape and 2 layers of black tape. Steps of 10min, 1cm. T=18oC. (Method described in run 3097) LY= 14.76 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3099 | 01/07/2016 | Marco | XP2262 (SN 34426) | SICCA J17 | -2.300 (NIM) | 149 (eff. 100) | LY scan | SICCA J17 crystal wrapped with ~5 layers of teflon tape and 2 layers of black tape. Bubbles placed on the opposite side of PMT window (closer to position 20cm in the plot). Steps of 10min, 1cm. T=18oC. (Method described in run 3097). Temperature not completely stable (1 hour only). LY= 25.26 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3100 | 01/08/2016 | Marco | XP2262 (SN 34426) | SICCA J17 | -2.300 (NIM) | 149 (eff. 100) | LY scan | Repeating run 3099, with stable temperature. LY= 24.28 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3115 (Entry$>4500) 3117..3120 |
01/15/2016 | Marco | XP4500 | (Aerogel 1.030) | -1.900 (NIM) | 149 | Aerogel prototype | Aerogel CUA's prototype to measure the light output differences in function of the aerogel refractive index. Prototype covered with alumidized mylar, 9 layers of aerogel, prototype aerogel top, 2 trigger PMTs in coincidence, artificial deadtime of 0.5us. Fit (range 0.5 to 10): 2.747pe. Mean (ch1>40 && ch1<800): 267.237 channels. |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3121 | 01/20/2016 | Marco | XP4500 | PMT calib | -1.900 (NIM) | 149 | PMT calib | Blue LED to calib SEP. SEP = 67.11 channels. PED = 27.03 channels. | Error creating thumbnail: Unable to save thumbnail to destination
| |
3122 | 01/20/2016 | Marco | XP4500 | PMT calib | -1.900 (NIM) | 149 | PMT calib | Repeating, trying to go with even lower light intensity. SEP = 64.30 channels. PED = 27.02 channels. Not sure if this is good enough, should try to reduce even more, but that is hard with the pulser we have. | Error creating thumbnail: Unable to save thumbnail to destination
| |
3124..3126 | 01/20/2016 | Marco | XP4500 | (Aerogel 1.011) | -1.900 (NIM) | 149 | Aerogel prototype | Prototype covered with alumidized mylar, 9 layers of aerogel, prototype aerogel top, 2 trigger PMTs in coincidence, artificial deadtime of 0.5us. Fit (range 0.5 to 3): 1.40011 . Mean (ch1>40 && ch1<800): 138.498 channels. |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3127..3132 | 01/27/2016 | Marco | XP4500 | (Aerogel 1.020) | -1.900 (NIM) | 149 | Aerogel prototype | Prototype covered with alumidized mylar, 9 layers of aerogel, prototype aerogel top, 2 trigger PMTs in coincidence, artificial deadtime of 0.5us. Tiles conditions not so good, scratches and chipped edges. Fit (range 0.5 to 5): 2.79186. Mean (ch1>40 && ch1<800): 203.788 channels. |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3133..3140 | 02/10/2016 | Marco | XP4500 | (Aerogel 1.015) | -1.900 (NIM) | 149 | Aerogel prototype | Prototype covered with alumidized mylar, 9 layers of aerogel, prototype aerogel top, 2 trigger PMTs in coincidence, artificial deadtime of 0.5us. Tiles conditions brand new. Fit (range 0.5 to 5): 2.39573. Mean (ch1>40 && ch1<800): 186.685 channels. |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3145 | 02/25/2016 | Marco | XP2262 (SN 34426) | SICCA J21 | -2.300 (NIM) | 100 | LY scan | SICCA J21 crystal wrapped with ~5 layers of teflon tape and 2 layers of black tape. Steps of 20min, 1cm. T=18oC. LY= 15.97 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3146 | 02/29/2016 | Marco | XP2262 (SN 34426) | SICCA J22 | -2.300 (NIM) | 100 | LY scan | SICCA J22. Same conditions as run 3145. Steps of 20min, 1cm. T=18oC. LY= 14.495 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3147 | 03/01/2016 | Marco | XP2262 (SN 34426) | SICCA J23 | -2.300 (NIM) | 100 | LY scan | SICCA J23. Same conditions as run 3145. Steps of 20min, 1cm. T=18oC. LY= 16.50 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3148 | 03/02/2016 | Marco | XP2262 (SN 34426) | SICCA J24 | -2.300 (NIM) | 100 | LY scan | SICCA J24. Same conditions as run 3145. Steps of 20min, 1cm. T=18oC. LY= 19.61 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3149 | 03/03/2016 | Marco | XP2262 (SN 34426) | SICCA J25 | -2.300 (NIM) | 100 | LY scan | SICCA J25. Same conditions as run 3145. Steps of 20min, 1cm. T=18oC. LY= 13.07 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3224 | 07/13/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | Prototype Optimization | Test to ensure that the prototype was functioning | Error creating thumbnail: Unable to save thumbnail to destination
| |
3225 | 07/14/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | Prototype Optimization | Test to ensure that the prototype was functioning | Error creating thumbnail: Unable to save thumbnail to destination
| |
3226 | 07/15/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | PMT Calibration | Finding the SEP for the PMT using a blue LED | Error creating thumbnail: Unable to save thumbnail to destination
| |
3227 | 07/15/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | PMT Calibration | Same setup as 3226 but with a smaller pulse width for the LED to ensure that the SEP was reached for this PMT | Error creating thumbnail: Unable to save thumbnail to destination
| |
3228 | 07/18/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | Aluminized Mylar Run | 3 day run to test the effectiveness of Aluminized Mylar | Error creating thumbnail: Unable to save thumbnail to destination
| |
3241 - 3259 | 07/26/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | Millipore Run | 4 day run to test the effectiveness of Millipore | Error creating thumbnail: Unable to save thumbnail to destination
| |
3270 - 3301 | 07/29/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | Teflon Tape Run | 2 day run to test the effectiveness of Teflon Tape | Error creating thumbnail: Unable to save thumbnail to destination
| |
3309 - 3327 | 08/01/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | Gore Run | 3 day run to test the effectiveness of Gore | Error creating thumbnail: Unable to save thumbnail to destination
| |
3335 - 3369 | 08/05/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | 9 Tiles | 3 Day run with 9 tiles of aerogel | Error creating thumbnail: Unable to save thumbnail to destination
| |
3370 - 3374 | 08/08/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | 8 Tiles | 3 day run with 8 tiles of aerogel | Error creating thumbnail: Unable to save thumbnail to destination
| |
3376 - 3397 | 08/10/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | 7 Tiles | 2 day run with 7 tiles of aerogel | Error creating thumbnail: Unable to save thumbnail to destination
| |
3400 - 34?? | 08/12/2016 | Salim | XP4500/B | - | 1800 (V) | 100 | 6 Tiles | 2 day run with 6 tiles of aerogel | File:3161.png | |
3421 | 11/16/2016 | Rich | XP2262 (SN 34426) | J19 20cm | -2.300 (NIM) | 100 | LY scan | J19 20 cm, T=18.0 oC. LY is a result of a gaussian fit of signal from crystal in the range of 170-800, SEP fit range 70-100, PED fit range 25-35; Gate width adjusted to get signal inside; 60K events LY= 14.04 pe/Mev |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3422 | 11/18/2016 | Rich | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 100 | LY scan | CRYTUR 10 cm, T=18.0 oC. LY is a result of a gaussian fit of signal from crystal in the range of 180-600, SEP fit range 60-110, PED fit range 25-35; Gate width adjusted to get signal inside; 100K events LY= 9.40 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3432 | 11/21/2016 | Salina | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 100 | LY scan | CRYTUR 10 cm, T=18.0 oC. LY is a result of a gaussian fit of signal from crystal in the range of 170-600, SEP fit range 60-110, PED fit range 25-35; Gate width adjusted to get signal inside; 100K events LY= 8.23 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3434 | 11/22/2016 | Salina | XP2262 (SN 34426) | CRYTUR 10cm | -2.300 (NIM) | 100 | LY scan | CRYTUR 10 cm, T=18.0 oC. LY is a result of a gaussian fit of signal from crystal in the range of 170-600, SEP fit range 65-100, PED fit range 25-35; Gate width same as 3432; 1.4M events LY= 9.67 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3445 | 12/06/2016 | Salina | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 100 | LY | J19, T=18.0 oC. LY is a result of a gaussian fit of signal from crystal in the range of 170-600 (Peak center: 309), SEP: Peak center @ 88.2 (fit range 65-100), PED: peak center @ 27.4 (fit range 25-35); Gate width adjusted from 80 ns to 100 ns; 160K events LY= 9.82 pe/MeV |
Error creating thumbnail: Unable to save thumbnail to destination
| |
3546 | 12/7/2016 | Rich/Salina | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 150 (100 effective) | LY | 100k, end unwrapped,10.03 pe/MeV | ||
3447-3470 | ??/??/???? | WHO? | WHAT? | WHAT? | WHAT? | ?? | ?? | Please ADD! If these are junk runs that's fine, but put some description here | ||
3470 | 12/12/2016 | Richard/TH | XP2262 (SN 34426) | J21 | -2.300 (NIM) | 100 | LY | J21 AFTER thermal annealing, T=18.0 oC. 12.09 pe/MeV | Temperature profile from thermal annealing of J21. We assume it is stable and was the same for J19. | |
3471 | 12/12/2016 | Richard/TH | XP2262 (SN 34426) | J21 | -2.300 (NIM) | 100 | LY | J21, T=18.0 oC. 11.55 pe/MeV | ||
3472 | 12/13/2016 | TH | XP2262 (SN 34426) | J21 | -2.300 (NIM) | 100 | LY | J21, T=18.0 oC. Same conditions as for runs 3470 and 3471. Quick analysis shows: LY= 12 pe/MeV | Temperature log for parts of 3471 and 3472 | |
3573 | 12/14/2016 | Rich | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 150 (100 effective) | LY | 100k, 14.90 pe/MeV | ||
3474 | 12/15/2016 | Marco/TH | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 100 | LY | J19, re-wrap the crystal. T~18.0 oC (not stable!!! just testing...) LY= 15.3 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3475 | 12/15/2016 | Marco/TH | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 100 | LY | J19, re-wrap the crystal. T~18.0 oC (1.5h since crystal was inserted) LY= pe/MeV | File:3475.png | |
3576 | 12/15/2016 | Rich | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 150 (100 effective) | LY | 100k, ~16 pe/MeV | ||
3577 | 12/15/2016 | Rich | XP2262 (SN 34426) | Crytur | -2.300 (NIM) | 150 (100 effective) | LY | 100k, ~16 pe/MeV | ||
3578 | 12/15/2016 | Rich | XP2262 (SN 34426) | J26 | -2.300 (NIM) | 150 (100 effective) | LY | 100k, 10.50 pe/MeV | ||
3579 | 12/15/2016 | Rich | XP2262 (SN 34426) | J36 | -2.300 (NIM) | 150 (100 effective) | LY | 100k, ~16 pe/MeV | ||
3481-3488 | 12/16/2016 | Andres | N/A | APDSetup | -2.300 (NIM) | 40 | APD Readout | Junk - APD Data rate acquisition test | ||
3489 | 12/19/2016 | TH | XP2262 (SN 34426) | J25 | -2.300 (NIM) | 150 (100 effective) | LY | J25 - 3 layer teflon, 2 layers black tape, centered on PMT, clamp loose, equilibrate for ~15 minutes, result: 12.5 pe/MeV (compare to 13.07 measured on 3/3/16) | ||
3490-3492 | 12/19/2016 | Andres | N/A | APDSetup | -2.300 (NIM) | 40 | APD Readout | Junk - ADC Test | ||
3493 | 12/19/2016 | TH | XP2262 (SN 34426) | J25 | -2.300 (NIM) | 150 (100 effective) | LY | J25 - 3 layer teflon, 2 layers black tape, centered on PMT, clamp loose, result: 12.7 pe/MeV (compare to 13.07 measured on 3/3/16) | ||
3494 | 12/19/2016 | TH | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY | J24 - 3 layer teflon, 2 layers black tape, centered on PMT, clamp loose, start run at 19.4 degC, result: 16.8 pe/MeV | ||
3495-3496 | 12/19/2016 | Andres | N/A | APDSetup | -2.300 (NIM) | 40 | APD Readout | Junk - ADC Test | ||
3497 | 12/19/2016 | TH | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY | J24 - 3 layer teflon, 2 layers black tape, centered on PMT, clamp loose, result: 16.7 pe/MeV | ||
3498 | 12/19/2016 | TH/Rishabh | XP2262 (SN 34426) | J36 | -2.300 (NIM) | 150 (100 effective) | LY | J36 - 3 layer teflon, 2 layers black tape, centered on PMT, clamp loose, equilibrate for ~15 minutes, result: 17 pe/MeV | ||
3499 | 12/19/2016 | TH/Rishabh | XP2262 (SN 34426) | J26 | -2.300 (NIM) | 150 (100 effective) | LY | J26 - 3 layer teflon, 2 layers black tape, centered on PMT, clamp loose, equilibrate for ~15 minutes, result:11.5 pe/MeV | ||
3500 | 12/20/2016 | TH | DAQ TEST | DAQ TEST | DAQ TEST | |||||
3501 | 12/20/2016 | TH | XP2262 (SN 34426) | J26 | -2.300 (NIM) | 150 (100 effective) | LY | J26 - same as run 3499, but after power outage over night on 12/20/16, result:11.95 pe/MeV | ||
3502 | 12/20/2016 | TH | XP2262 (SN 34426) | CRYTUR01 | -2.300 (NIM) | 150 (100 effective) | LY | CRYTUR01 - source aligned at the top (did not move it from position used for 20cm crystal), start run at 20 degC, result:15.5 pe/MeV | ||
3503 | 12/20/2016 | TH | XP2262 (SN 34426) | CRYTUR01 | -2.300 (NIM) | 150 (100 effective) | LY | CRYTUR01 - same as run 3502, note that at end of run gate was briefly removed, result:15.7 pe/MeV | ||
3504 | 12/20/2016 | TH | XP2262 (SN 34426) | CRYTUR01 | -2.300 (NIM) | 150 (100 effective) | LY | CRYTUR01 - same as run 3503, but with the other end of crystal taped shut, result:14.8 pe/MeV | ||
3505 | 12/20/2016 | TH | XP2262 (SN 34426) | CRYTUR01 | -2.300 (NIM) | 150 (100 effective) | LY | CRYTUR01 - same as run 3504, but with source aligned at center of crystal using stepper motor, result:15.7 pe/MeV | ||
3506 | 12/20/2016 | TH | XP2262 (SN 34426) | J17 | -2.300 (NIM) | 150 (100 effective) | LY | J17 - source aligned at center of crystal using stepper motor, this is the crystal with bubbles inside, result:26 pe/MeV | ||
3507 | 12/20/2016 | TH | XP2262 (SN 34426) | J18 | -2.300 (NIM) | 150 (100 effective) | LY | J18 - same as run 3506, result:11.3 pe/MeV | ||
3508 | 12/20/2016 | TH | XP2262 (SN 34426) | J31 | -2.300 (NIM) | 150 (100 effective) | LY | J31 - long run overnight, same conditions as 3507, result: 13.64 pe/MeV | ||
3509 | 12/21/2016 | TH | XP2262 (SN 34426) | J39 | -2.300 (NIM) | 150 (100 effective) | LY | J39 - same conditions as 3508, start run at 19.6 degC, result: 11.9 pe/MeV | ||
3510 | 12/21/2016 | TH | XP2262 (SN 34426) | J44 | -2.300 (NIM) | 150 (100 effective) | LY | J44 - same conditions as 3509, result: 12.4 pe/MeV | ||
3511 | 12/21/2016 | TH | XP2262 (SN 34426) | J15 | -2.300 (NIM) | 150 (100 effective) | LY | J15 - same conditions as 3510, this is a crystal with a chip on one end, for this measurement the chipped end is taped shut, result:XX pe/MeV | ||
3512 | 1/3/2017 | TH | XP2262 (SN 34426) | J24/Reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - first run after holiday shutdown, re-wrapped crystal, experimental conditions same as in December runs, start run at 19. degC, result: 16.3 pe/MeV | ||
3513-3521 | 1/3/2017 | TH | N/A | N/A | N/A | N/A | LY | Junk - DAQ configuration tests | ||
3521 | 1/3/2017 | TH | XP2262 (SN 34426) | J24/Reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - experimental conditions same as in 3521, result:19.02 pe/MeV | ||
3522 | 1/3/2017 | TH | XP2262 (SN 34426) | J24/Reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - experimental conditions same as in 3521, result:17 pe/MeV (~140k stats, after ~40k stats the fits with same limits gave 18.7 pe/MeV) | ||
3523 | 1/3/2017 | TH | XP2262 (SN 34426) | J24/Reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - long run overnight, experimental conditions same as in 3522, result: 17.6 pe/MeV | ||
3524 | 1/4/2017 | TH | XP2262 (SN 34426) | J24/Reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - experimental conditions same as in 3523, result:17.9 pe/MeV (~100k stats, after ~50k stats the fits with same limits gave 18 pe/MeV); SEP shifted by about 1.5 channels compared to 3523, signal shifted by about 5 channels in the same direction as the SEP | ||
3525 | 1/4/2017 | TH | XP2262 (SN 34426) | J24/Reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - remove optical grease on both crystal and PMT, re-apply optical grease and re-install crystal, misalign crystal on purpose on PMT towards the right away from center as viewed from the trigger scintillator, result:16.3 pe/MeV (~230k stats), final result: 16.2 pe/MeV | ||
3526 | 1/4/2017 | TH | XP2262 (SN 34426) | J24/Reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - same as 3525, but better thermally equilibrated, result: 17.3 pe/MeV (~230k stats) | ||
3527 | 1/4/2017 | TH | XP2262 (SN 34426) | J39 | -2.300 (NIM) | 150 (100 effective) | LY | J39 - aligned crystal to center of PMT, start run at 19degC, result: 12.25 pe/MeV | ||
3528 | 1/4/2017 | TH | XP2262 (SN 34426) | J39 | -2.300 (NIM) | 150 (100 effective) | LY | J39 - same as 3527, note that with same fitting limits the fit to SEP is not very good, result: 11.9 pe/MeV | ||
3529 | 1/4/2017 | TH | XP2262 (SN 34426) | J39 | -2.300 (NIM) | 150 (100 effective) | LY | J39 - same as 3527, not enough statistics to fit | ||
3530 | 1/4/2017 | TH | XP2262 (SN 34426) | J36 | -2.300 (NIM) | 150 (100 effective) | LY | J36 - randomly selected for long overnight run, conditions same as for 3529, result: 17.17 pe/MeV | ||
3531 | 1/5/2017 | TH | XP2262 (SN 34426) | J36 | -2.300 (NIM) | 150 (100 effective) | LY | J36 - same as 3530, result: 16.54 pe/MeV | ||
3532 | 1/4/2017 | TH | XP2262 (SN 34426) | J16 | -2.300 (NIM) | 150 (100 effective) | LY | J16 - already wrapped, was irradiated and LED cured in summer 2016, one of the worst crystals for radiation hardness based on summer 16, result: 9.68 pe/MeV | ||
3533 | 1/4/2017 | TH | XP2262 (SN 34426) | J16 | -2.300 (NIM) | 150 (100 effective) | LY | J16 - same as 3532, result: 9.83 pe/MeV | ||
3534 | 1/5/2017 | TH | XP2262 (SN 34426) | J23 | -2.300 (NIM) | 150 (100 effective) | LY | J23 - wrapped this one (3 layers teflon and 2 layers black tape - that's how I wrapped all crystals I tested), this crystal was also irradiated in and LED cured in summer 2016, the lowest radiation hardness based on summer 16, result: 8.07 pe/MeV | ||
3535 | 1/5/2017 | TH | XP2262 (SN 34426) | J21 | -2.300 (NIM) | 150 (100 effective) | LY | J21 - already wrapped, was irradiated and LED cured in summer 2016, thermally annealed in December 2016, one of the worst crystals for radiation hardness based on summer 16, result: 12.94 pe/MeV | ||
3536 | 1/5/2017 | TH | XP2262 (SN 34426) | J21 | -2.300 (NIM) | 150 (100 effective) | LY | J21 - same as 3535, result: 12.81 pe/MeV | ||
3537 | 1/5/2017 | TH | XP2262 (SN 34426) | J24/reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - same conditions as 3536, result: 17.42 pe/MeV - consistent with run 3526 | ||
3538 | 1/6/2017 | Rich/TH | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY | J32 - Rich installs this one and all crystals up to run 3543 (we may have a different technique - needs to be checked), result: 11.93 pe/MeV | ||
3539 | 1/6/2017 | Rich/TH | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY | J32 - same as 3538, result: 13.05 pe/MeV, note: irradiated crystal after this measurement; another note: crystal accident during transmittance measurement chips off at two places off one end of crystal | ||
3540 | 1/5/2017 | Rich/TH | XP2262 (SN 34426) | J38 | -2.300 (NIM) | 150 (100 effective) | LY | J38 - same conditions as 3539, result: 12.0 pe/MeV | ||
3541 | 1/6/2017 | Rich/TH | XP2262 (SN 34426) | J27 | -2.300 (NIM) | 150 (100 effective) | LY | J32 - Rich installs this one and all crystals up to run 3543 (we may have a different technique - needs to be checked), result: 11.93 pe/MeV, note: irradiated crystal after this measurement | ||
3542 | 1/6/2017 | Rich/TH | XP2262 (SN 34426) | J24/reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - same as 3541, result: 12.79 pe/MeV | ||
3543-45 | 1/6/2017 | Rich/TH | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY | J24 - to test crystal installation technique, TH applies optical grease, installs crystal etc. for this run, no signal from crystal in these runs as it had tilted off PMT and caught on arm that holds the source, no obvious damage to crystal | ||
3546 | 1/6/2017 | Rich/TH | XP2262 (SN 34426) | J24/reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - to test crystal installation technique, TH applies optical grease, installs crystal etc. for this run, fixed temperature sensor cable back to freezer wall - it had come loose and was partially covering the scintillator, we also noticed that the crystal PMT seems more loose now - perhaps slightly tilted (need to check), crystal is positioned a little more to left from center as viewed from scintillator PMT, we used the clamp to hold crystal in this run to restrict motion, after a few events we verified that we got signal from crystal now, long run over weekend, result: 16.99 +-0.05 pe/MeV | ||
3547 | 1/9/2017 | TH | XP2262 (SN 34426) | J24/reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - result: 16.68 +-0.41 pe/MeV | ||
3548 | 1/9/2017 | TH | XP2262 (SN 34426) | J40 | -2.300 (NIM) | 150 (100 effective) | LY | J40 - this crystal has relatively large broken off pieces on both ends, I taped shut the side that has a chip that seems to alter the crystal surface, 3 teflon layers and 2 black tape as usual, install with clamp, PMT seems still a little tilted, re-taped temperature sensor cable to freezer wall, result: 13.32 pe/MeV | ||
3549 | 1/9/2017 | TH | XP2262 (SN 34426) | J40 | -2.300 (NIM) | 150 (100 effective) | LY | J40 - like 3548 but better thermally equilibrated, result: 14 pe/MeV | ||
3550 | 1/9/2017 | TH | XP2262 (SN 34426) | J38 | -2.300 (NIM) | 150 (100 effective) | LY | J38 - already wrapped (1 layer of teflon and 1 layer black tape as find out later), tighten crystal PMT holder before this run and also re-adjust clamp, result: 12.21 pe/MeV | ||
3551 | 1/9/2017 | TH | XP2262 (SN 34426) | J38 | -2.300 (NIM) | 150 (100 effective) | LY | J38 - same as 3550, result: 13.39 pe/MeV | ||
3552 | 1/9/2017 | TH | XP2262 (SN 34426) | J38 | -2.300 (NIM) | 150 (100 effective) | LY | J38 - same as 3552, result: 12.6 pe/MeV | ||
3553 | 1/9/2017 | TH | XP2262 (SN 34426) | J27 | -2.300 (NIM) | 150 (100 effective) | LY | J27 - already wrapped, re-attach temperature sensor cable to freezer wall, result: 14.3 pe/MeV | ||
3554 | 1/9/2017 | TH | XP2262 (SN 34426) | J27 | -2.300 (NIM) | 150 (100 effective) | LY | J27 - same as 3553, long overnight run, result: 14.16 pe/MeV | ||
3555 | 1/10/2017 | TH | XP2262 (SN 34426) | J27 | -2.300 (NIM) | 150 (100 effective) | LY | J27 - same as 3554, result: 13.78 pe/MeV | ||
3556 | 1/10/2017 | TH | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY | J32 - was irradiated on 6 January 2017 transverse, re-wrap crystal (3 teflon, 2 black tape layers), I taped shut the side with the chip after our 6 January 2017 accident, start data taking at 19 degC, result: 14.51 pe/MeV | ||
3557 | 1/10/2017 | TH | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY | J32 - same as 3556, but better thermally equilibrated, result: 16.26 pe/MeV | ||
3558 | 1/10/2017 | TH | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY | J32 - same as 3557, result: 15.56 pe/MeV | ||
3559 | 1/10/2017 | TH | XP2262 (SN 34426) | J45 | -2.300 (NIM) | 150 (100 effective) | LY | J45 - already wrapped, open side has brownish marks that cannot be removed with alcohol (observed similar marks on J32), start data taking at 19.5degC, result: 14.02 pe/MeV | ||
3560 | 1/10/2017 | TH | XP2262 (SN 34426) | J45 | -2.300 (NIM) | 150 (100 effective) | LY | J45 - same as 3559, but better thermally equilibrated, result: 13.41 pe/MeV | ||
3561 | 1/10/2017 | TH | XP2262 (SN 34426) | J31 | -2.300 (NIM) | 150 (100 effective) | LY | J31 - already wrapped, start run at 18.7degC, result: 15.38 pe/MeV | ||
3562 | 1/10/2017 | TH | XP2262 (SN 34426) | J31 | -2.300 (NIM) | 150 (100 effective) | LY | J31 - long overnight run, result: 15.29 pe/MeV | ||
3563 | 1/11/2017 | TH | XP2262 (SN 34426) | J31 | -2.300 (NIM) | 150 (100 effective) | LY | J31 - same as 3562, result: 15.11 pe/MeV | ||
3564 | 1/11/2017 | TH | XP2262 (SN 34426) | J35 | -2.300 (NIM) | 150 (100 effective) | LY | J35 - already wrapped, just taped one end, result: 11.64 pe/MeV (note: not a good fit to SEP crystal with the current limits in LY.C - need more careful analysis) | ||
3565 | 1/11/2017 | TH | XP2262 (SN 34426) | J35 | -2.300 (NIM) | 150 (100 effective) | LY | J35 - like 3564, result: 12.37 pe/MeV (note: not a good fit to SEP crystal with the current limits in LY.C - need more careful analysis) | ||
3566 | 1/11/2017 | TH | XP2262 (SN 34426) | J35 | -2.300 (NIM) | 150 (100 effective) | LY | J35 - like 3565, low stats run, result: 13.79 pe/MeV (note: not a good fit to SEP crystal with the current limits in LY.C - need more careful analysis) | ||
3567 | 1/11/2017 | TH | XP2262 (SN 34426) | J11 | -2.300 (NIM) | 150 (100 effective) | LY | J11 - this crystal was tested previously at Caltech and was determined to have low radiation hardness and many scattering centers, result: 9.81 pe/MeV (Caltech LY result: 12.5 pe/MeV) | ||
3568 | 1/11/2017 | TH | XP2262 (SN 34426) | J11 | -2.300 (NIM) | 150 (100 effective) | LY | J11 -same as 3567, result: 10.25 pe/MeV (Caltech LY result: 12.5 pe/MeV) | ||
3569 | 1/12/2017 | TH | XP2262 (SN 34426) | J11 | -2.300 (NIM) | 150 (100 effective) | LY | J11 -same as 3568, result: 10.31 pe/MeV (Caltech LY result: 12.5 pe/MeV) | ||
3570 | 1/12/2017 | TH | XP2262 (SN 34426) | J33 | -2.300 (NIM) | 150 (100 effective) | LY | J33 - already wrapped, only taped shut one end, start run at 18.9 degC, result: 11.74 pe/MeV | ||
3571 | 1/12/2017 | TH | XP2262 (SN 34426) | J33 | -2.300 (NIM) | 150 (100 effective) | LY | J33 - like 3570, result: 12.26 pe/MeV | ||
3572 | 1/12/2017 | TH | XP2262 (SN 34426) | J34 | -2.300 (NIM) | 150 (100 effective) | LY | J34 - already wrapped, start run at 19degC, result: 10.3 pe/MeV | ||
3573 | 1/12/2017 | TH | XP2262 (SN 34426) | J34 | -2.300 (NIM) | 150 (100 effective) | LY | J34 - like 3572, result: 9.4 pe/MeV | ||
3574 | 1/12/2017 | TH | XP2262 (SN 34426) | J24/reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - long overnight run, result: 16.93 +-0.12 pe/MeV | ||
3575 | 1/13/2017 | TH | XP2262 (SN 34426) | J24/reference | -2.300 (NIM) | 150 (100 effective) | LY | J24 - same as 3574, result: | ||
3577 | 1/13/2017 | Rich | XP2262 (SN 34426) | J34 | -2.300 (NIM) | 150 (100 effective) | LY | Significant amount of optical grease, 30 min, ~10.3 pe/MeV | ||
3578 | 1/13/2017 | Rich | XP2262 (SN 34426) | J34 | -2.300 (NIM) | 150 (100 effective) | LY | Very little optical grease, 30 min, no signal | ||
3579 | 1/13/2017 | Rich | XP2262 (SN 34426) | J34 | -2.300 (NIM) | 150 (100 effective) | LY | No run | ||
3580 | 1/13/2017 | Rich | XP2262 (SN 34426) | J34 | -2.300 (NIM) | 150 (100 effective) | LY | Very little optical grease, 30 min, ~9.8 pe/MeV | ||
3581 | 1/13/2017 | Rich | XP2262 (SN 34426) | J24/reference | -2.300 (NIM) | 150 (100 effective) | LY | Significant amount of optical grease, 30 min, ~8.9 pe/MeV | ||
3582 | 1/13/2017 | Rich | XP2262 (SN 34426) | J24/reference | -2.300 (NIM) | 150 (100 effective) | LY | Very little optical grease, 30 min, ~5.23 pe/MeV | ||
3583 | 1/13/2017 | Rich | XP2262 (SN 34426) | J24/reference | -2.300 (NIM) | 150 (100 effective) | LY | Pea size of optical grease, 30 min, ~15.9 pe/MeV. The variation of optical grease over the last couple of runs (3577-83) results in fluctuating light yields, a more consistent method should be developed | ||
3591 | 1/31/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 18°C | Crystal equilibrated start @ 14:11; t=0 end at t=64 min, 11.26 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3592 | 1/31/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 18°C | Crystal equilibrated (continuing); t=65 min, end at t=125 min, 11.71 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3593 | 1/31/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 18°C | Crystal equilibrated (continuing) t=126 min, end at t=186 min, 11.54 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3594 | 1/31/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 18°C | Crystal equilibrated (continuing with +149 min delay) @ start time: 19:47; t=335 min, end at t= 695 min, 11.97 pe/MeV; 6 hour/360 min run | Error creating thumbnail: Unable to save thumbnail to destination
| |
3595 | 2/1/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 15°C | Freezer opened for ~ 2 min to check dehumidifier; Started run after 2 hours. t=0 min, end at t=60 min, 12.64 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3596 | 2/1/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 15°C | Crystal equilibrated t=77 min, end at t=137 min, 12.72 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3597 | 2/1/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 15°C | Crystal equilibrated t=167 min, end at t= 227 min, 12.56 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3598 | 2/1/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 15°C | Crystal equilibrated t=228 min, end at t= 588 min, 12.67 pe/MeV; 6 hour/360 min run | Error creating thumbnail: Unable to save thumbnail to destination
| |
3599 | 2/3/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 10°C | Crystal equilibrated for 24 hours; taking 1 hour runs , 14.82 pe/MeV; | Error creating thumbnail: Unable to save thumbnail to destination
| |
3600 | 2/3/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 10°C | Taking 1 hour runs , 14.14 pe/MeV; | Error creating thumbnail: Unable to save thumbnail to destination
| |
3601 | 2/3/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 10°C | Taking 1 hour runs , 14.68 pe/MeV; | Error creating thumbnail: Unable to save thumbnail to destination
| |
3602 | 2/5/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 10°C | 6 hour/360 min run , 14.85 pe/MeV; | Error creating thumbnail: Unable to save thumbnail to destination
| |
3603 | 2/5/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T dependence of LY | Junk run..only took data for 3 min; checking temperature | ||
3604 | 2/6/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 5°C | Crystal equilibrated for 120 min, after opening freezer for a couple of minutes. Taking 1 hour runs , 17.13 pe/MeV; | Error creating thumbnail: Unable to save thumbnail to destination
| |
3605 | 2/6/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 5°C | Crystal equilibrated t=70 min end at t= 130. Taking 1 hour runs , 17.04 pe/MeV; | Error creating thumbnail: Unable to save thumbnail to destination
| |
3606 | 2/6/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 5°C | Crystal equilibrated t=190 min, end at t= 250 min, 18.05 pe/MeV; Taking 1 hour runs | Error creating thumbnail: Unable to save thumbnail to destination
| |
3607 | 2/6/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 5°C | Crystal equilibrated t=265 min, end at t= 625min, 17.25 pe/MeV; 6 hour/360 min run; Note: need to look at temperature as function of time in freezer for this run. | Error creating thumbnail: Unable to save thumbnail to destination
| |
3609 | 2/9/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 5°C | Crystal equilibrated >24 hours, 17.15 pe/MeV; 1 hour run | Error creating thumbnail: Unable to save thumbnail to destination
| |
3610 | 2/10/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at 0°C | Crystal equilibrated for > 24 hours at 0°C after the freezer was left open (2/7/2017). 21.86 pe/MeV; Taking 1 hour runs, and will not open freezer until after runs are completed to avoid moisture buildup in freezer. | Error creating thumbnail: Unable to save thumbnail to destination
| |
3611 | 2/11/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at -5°C | Crystal equilibrated for > 24 hours at -5°C. 23.6 pe/MeV; 1 hour | Error creating thumbnail: Unable to save thumbnail to destination
| |
3612 | 2/13/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at -10°C | Crystal equilibrated for > 24 hours at -10°C. 29.16 pe/MeV; 1 hour run. | Error creating thumbnail: Unable to save thumbnail to destination
| |
3613 | 2/14/2017 | SA | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | T Dependence of LY at -10°C | Additional data for statistics @ -10°C. 30.37 pe/MeV; 1 hour run. | Error creating thumbnail: Unable to save thumbnail to destination
| |
3636 | 5/17/2017 | RT | XP2262 (SN 34426) | J24 (reference) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | Check reference before Crytur II run @ 18°C. 15.8 pe/MeV; 1 hour run | -- | |
3637 | 5/17/2017 | RT | XP2262 (SN 34426) | Cytur II | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | First runs for Crytur II @ 18°C. 10.25 pe/MeV; May not have reached equilibrium; 30 min. run | -- | |
3638 | 5/17/2017 | RT | XP2262 (SN 34426) | Cytur II | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | First runs for Crytur II @ 18°C. 10.4 pe/MeV; 30 min. run | -- | |
3639 | 5/17/2017 | RT | XP2262 (SN 34426) | Cytur II | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | First runs for Crytur II @ 18°C. Grease reapplied. 11.4 pe/MeV; Overnight run | -- | |
3642 | 5/31/2017 | Marco/Tanja | XP2262 (SN 34426) | Cytur II | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | Same as before, running overnight. 15.19 pe/MeV - Double peak observed in pedestal. Is that related to the APD cooler that is running in the lab? To investigate... | Error creating thumbnail: Unable to save thumbnail to destination
| |
3647 | 6/1/2017 | RT | XP2262 (SN 34426) | Cytur II | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | Ran for an hour. 11.97 pe/MeV - Double peak observed in pedestal. Need too investigate... | -- | |
3648 | 6/1/2017 | RT | XP2262 (SN 34426) | Cytur II | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | Ran for an hour. 10.57 pe/MeV - Double peak observed in pedestal. Need too investigate... | -- | |
3671 | 7/5/2017 | RT | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | Quick run 15 min. 13.8 pe/MeV - Double peak observed in pedestal. The first test before trying to figure out this double peak. | -- | |
3674-3676 | 7/5/2017 | RT | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | Various double peak while changing Andres set up to see any effects. There were none. | -- | |
3677 | 7/5/2017 | RT | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | Used the pulse generator at 150 ns gate width to test double peak. There was no visible double peak but there is a noticeable smaller spike coming from the scintillator gate. | Error creating thumbnail: Unable to save thumbnail to destination
| |
3678-3687 | 7/5/2017 | RT | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | Various double peak while changing Andres set up and changing gate widths to see any effects. There were none. | -- | |
3689 | 7/10/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. result:14.056 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3690 | 7/10/2017 | Summer Students | XP2262 (SN 34426) | J16 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. result:8.798 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3691 | 7/10/2017 | Summer Students | XP2262 (SN 34426) | J18 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. result:10.26 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3692 | 7/10/2017 | Summer Students | XP2262 (SN 34426) | J21 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. result:10.89 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3693 | 7/10/2017 | Summer Students | XP2262 (SN 34426) | J27 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. result:11.89 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3694 | 7/10/2017 | Summer Students | XP2262 (SN 34426) | J41 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. result: 7.587 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3695 | 7/10/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. result:10.97 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3698 | 7/11/2017 | Summer Students | XP2262 (SN 34426) | J41 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. result:8.83897 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3707 | 7/11/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Wrapped with one layer ESR, and two layers of black tape. Double peak observed in pedestal. result:11.22 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3709 | 7/11/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. result:10.5329 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3713 | 7/11/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Many peaks observed in pedestal. Most likely voltage or gate was changed during the trial. result:11.83 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3722 | 7/14/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result:14.51 +/- .82 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3723 | 7/14/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. Crystal was left in freezer with optical for several days, possible causing the low light yield. result:5.68889 +/- 2.47504 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3725 | 7/14/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer ESR material, 2 layers electrical tape. result:13.1433 +/- .811166 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3726 | 7/14/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer Mylar Foil material, 2 layers electrical tape. result: 8.98522 +/- 3.06839 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3727 | 7/14/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 19.4425 +/- 11.9027 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3728 | 7/17/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 11.32 +/- .49 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3729 | 7/17/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 11.41 +/- .44 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3730 | 7/17/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer ESR reflector, 2 layers electrical tape. result: 11.83 +/- .49 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3731 | 7/17/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer Tedlar reflector, 2 layers electrical tape. result: 5.45 +/- .29 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3732 | 7/17/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 14.80 +/- .35 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3733 | 7/17/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 11.62 +/- .46 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3734 | 7/18/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 12.93 +/- .63 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3736 | 7/18/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 14.36 +/- .70 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3737 | 7/18/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 14.80 +/- .74 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3738 | 7/18/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layers teflon tape, 2 layers electrical tape. result: 8.58 +/- .30 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3739 | 7/18/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer Mylar Foil reflector, 2 layers electrical tape. result: 9.05 +/- .36 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3740 | 7/18/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer ESR reflector, 2 layers electrical tape. result: 13.22 +/- .59 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3741 | 7/18/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layers teflon tape, 2 layers electrical tape. result: 8.86 +/- .38 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3742 | 7/18/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer Mylar Foil reflector, 2 layers electrical tape. result: 10.57 +/- .36 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3743 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer ESR material, 2 layers electrical tape. result: 15.83 +/- .68 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3744 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layers teflon tape, 2 layers electrical tape. result: 7.83 +/- .37 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3745 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layers teflon tape, 2 layers electrical tape. result: 8.10 +/- .59 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3746 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. top half wrapped in Tedlar, bottom half 3 layers teflon tape, 2 layers electrical tape. result: 6.62 +/- .42 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3747 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 2 layers teflon tape, 2 layers electrical tape. result: 10.84 +/- .63 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3748 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. top half wrapped in Tedlar, bottom half 3 layers teflon tape, 2 layers electrical tape. result: 6.58 +/- .51 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3749 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 2 layers teflon tape, 2 layers electrical tape. result: 10.25 +/- .29 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3750 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 2 layer Teflon Tape reflector, 2 layers electrical tape. result: 11.42 +/- .64 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3751 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. top half wrapped in 3 layers teflon tape, bottom half tedlar, 2 layers electrical tape. result: 10.24 +/- .30 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3752 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. top half wrapped in 3 layers teflon tape, bottom half tedlar, 2 layers electrical tape. result: 10.57 +/- .48 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3753 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. op half wrapped in 3 layers teflon tape, bottom half tedlar, 2 layers electrical tape. bubbles in optical grease resulting in error, result: 8.18 +/- .72 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3754 | 7/19/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. bubbles in optical grease resulting in error, result: 8.49 +/- .93 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3755 | 7/20/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 11.13 +/- .74 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3756 | 7/20/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. top half wrapped in Tedlar, bottom half 3 layers teflon tape, 2 layers electrical tape. result: 6.93 +/- .25 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3757 | 7/20/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 11.00 +/- .35 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3758 | 7/21/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 11.34 +/- .46 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3759 | 7/21/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. top half wrapped in 3 layers teflon tape, bottom half tedlar, 2 layers electrical tape. result: 9.44 +/- .33 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3760 | 7/21/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 4 layer Teflon Tape reflector, 2 layers electrical tape. result: 11.97 +/- .46 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3761 | 7/21/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 4 layer Teflon Tape reflector, 2 layers electrical tape. result: 12.24 +/- .52 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3762 | 7/21/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 4 layer Teflon Tape reflector, 2 layers electrical tape. result: 12.48 +/- .89 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3763 | 7/21/2017 | Summer Students | XP2262 (SN 34426) | J24 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. top half wrapped in 3 layers teflon tape, bottom half tedlar, 2 layers electrical tape. result: 9.11 +/- .49 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3764 | 7/21/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, end wrapped in Tedlar, 2 layers electrical tape. result: 8.75 +/- .27 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3765 | 7/21/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, end wrapped in Tedlar, 2 layers electrical tape. result: 9.38 +/- .42 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3766 | 7/21/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, end wrapped in Tedlar, 2 layers electrical tape. result: 9.22 +/- .56 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3768 | 7/24/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 5 layers teflon tape, 2 layers electrical tape. result: 12.88 +/- .59 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3769 | 7/24/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 5 layers teflon tape, 2 layers electrical tape. result: 11.50 +/- .53 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3770 | 7/24/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 5 layer Teflon Tape reflector, 2 layers electrical tape. result: 11.63 +/- .39 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3771 | 7/25/2017 | Summer Students | XP2262 (SN 34426) | J42 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 5 layer Teflon Tape reflector, 2 layers electrical tape. result: 10.81 +/- .31 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3772 | 8/1/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layer Teflon Tape reflector, 2 layers electrical tape. result: 10.02 +/- .45 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3773 | 8/1/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layer Teflon Tape reflector, 2 layers electrical tape. result: 12.17 +/- .58 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3774 | 8/1/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 12.85 +/- .55 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3775 | 8/1/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 11.33 +/- .32 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3776 | 8/2/2017 | Summer Students | XP2262 (SN 34426) | J32 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 11.21 +/- .69 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3777 | 8/2/2017 | Summer Students | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 11.80 +/- .46 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3778 | 8/2/2017 | Summer Students | XP2262 (SN 34426) | J35 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 13.37 +/- .50 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3779 | 8/2/2017 | Summer Students | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 11.93 +/- .46 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3780 | 8/2/2017 | Summer Students | XP2262 (SN 34426) | J35 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layer Teflon Tape reflector, 2 layers electrical tape. result: 12.34 +/- .39 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3782 | 8/2/2017 | Summer Students | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer gortex reflector, 2 layers electrical tape. result: 16.59 +/- .60 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3783 | 8/2/2017 | Summer Students | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer gortex reflector, 2 layers electrical tape. result: 16.45 +/- .45 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3784 | 8/2/2017 | Summer Students | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer gortex reflector, 2 layers electrical tape. result: 16.42 +/- .60 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3785 | 8/3/2017 | Summer Students | XP2262 (SN 34426) | J35 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 3 layers teflon tape, 2 layers electrical tape. result: 12.34 +/- .33 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3786 | 8/3/2017 | Summer Students | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer 200 micrometer Mylar, 2 layers electrical tape. result: 8.96 +/- .21 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3787 | 8/3/2017 | Summer Students | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer 200 micrometer Mylar, 2 layers electrical tape. result: 8.68 +/- .34 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3788 | 8/3/2017 | Summer Students | XP2262 (SN 34426) | J19 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | ~50,000 events - Double peak observed in pedestal. wrapped in 1 layer 200 micrometer Mylar, 2 layers electrical tape. result: 9.10 +/- .23 pe/MeV | Error creating thumbnail: Unable to save thumbnail to destination
| |
3789 | 9/18/2017 | Tanja | XP2262 (SN 34426) | CRYTUR82 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | With crystal holder, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 11.2 pe/MeV | ||
3790 | 9/19/2017 | Tanja | XP2262 (SN 34426) | CRYTUR82 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | With crystal holder, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 11.8 pe/MeV | ||
3791 | 9/19/2017 | Tanja | XP2262 (SN 34426) | J24 (main reference) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | With crystal holder, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 14.3 +- 0.6 pe/MeV | ||
3792 | 9/19/2017 | Tanja | XP2262 (SN 34426) | J24 (main reference) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | With crystal holder, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 15.1 +- 0.3 pe/MeV | ||
3793 | 9/19/2017 | Tanja | XP2262 (SN 34426) | J24 (main reference) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | With crystal holder, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: ? pe/MeV | ||
3794 | 9/19/2017 | Tanja | XP2262 (SN 34426) | J39 (reference #2) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | With crystal holder, double peak observed in pedestal. Was already wrapped (looks like 1 layer teflon and one layer black tape). result: 10 +- 0.2 pe/MeV | ||
3795 | 9/19/2017 | Tanja | XP2262 (SN 34426) | J39 (reference #2) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | With crystal holder, double peak observed in pedestal. Was already wrapped (looks like 1 layer teflon and one layer black tape). result: 11 +- 0.5 pe/MeV | ||
3796 | 9/19/2017 | Tanja | XP2262 (SN 34426) | J39 (reference #2) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | Like previous, but WITHOUT crystal holder using a clamp, double peak observed in pedestal. Was already wrapped (looks like 1 layer teflon and one layer black tape). result: 13 +- 0.4 pe/MeV | ||
3797 | 9/19/2017 | Tanja | XP2262 (SN 34426) | J39 (reference #2) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | Like previous, but WITHOUT crystal holder using a clamp, double peak observed in pedestal. Was already wrapped (looks like 1 layer teflon and one layer black tape). result: 13.2 +- 0.1 pe/MeV | ||
3798 | 9/22/2017 | Tanja | XP2262 (SN 34426) | J24 (main reference) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | After Thermal Annealing, WITHOUT crystal holder using a clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 15.8 +- 0.4 pe/MeV | ||
3799 | 9/22/2017 | Tanja | XP2262 (SN 34426) | J24 (main reference) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | After Thermal Annealing, WITHOUT crystal holder using a clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 16.4 +- 0.4 pe/MeV | ||
3800 | 9/22/2017 | Tanja | XP2262 (SN 34426) | J39 (reference #2) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | After Thermal Annealing, WITHOUT crystal holder using a clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 10.7 +- 0.3 pe/MeV (100k events) | ||
3801 | 9/22/2017 | Tanja | XP2262 (SN 34426) | J39 (reference #2) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | After Thermal Annealing, WITHOUT crystal holder using a clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 10.8 +- 0.2 pe/MeV (100k events) | ||
3802 | 9/22/2017 | Tanja | XP2262 (SN 34426) | CRYTUR #82 | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | After Thermal Annealing, WITHOUT crystal holder using a clamp, tape shut the end with bigger crack, equilibrate about 20 minutes (shorter than 3798-3801), double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 15.1 +- 0.6 pe/MeV (100k events) | ||
3803 | 10/3/2017 | Tanja | Junk - DAQ rebooting VME1 with minicom | |||||||
3804 | 10/3/2017 | Tanja | XP2262 (SN 34426) | J24 (main reference) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | After one-week shutdown and DAQ reboot, no other changes from 3799, WITHOUT crystal holder using a clamp, equilibrate ~40 min., double peak observed in pedestal. Still wrapped in 3 layers teflon, 2 layers black tape. result: 17.2 +- 0.8 pe/MeV | ||
3805 | 10/3/2017 | Tanja | XP2262 (SN 34426) | J24 (main reference) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | using clamp, long run, double peak observed in pedestal. Still wrapped in 3 layers teflon, 2 layers black tape. result: 16.2 +- 0.1 pe/MeV | ||
3806 | 10/4/2017 | Tanja | XP2262 (SN 34426) | J39 (reference #2) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | using clamp, equilibrate 30min, double peak observed in pedestal. Still wrapped in 3 layers teflon, 2 layers black tape. result: 11.1 +- 0.3 pe/MeV | ||
3807 | 10/4/2017 | Tanja | XP2262 (SN 34426) | J39 (reference #2) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | using clamp, double peak observed in pedestal. Still wrapped in 3 layers teflon, 2 layers black tape. result: 12.1 +- 0.8 pe/MeV | ||
3808 | 10/4/2017 | Tanja | XP2262 (SN 34426) | J25 (reference #3) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | using clamp, double peak observed in pedestal. Still wrapped in 3 layers teflon, 2 layers black tape. result: 12.0 +- 0.4 pe/MeV | ||
3809 | 10/4/2017 | Tanja | XP2262 (SN 34426) | J25 (reference #3) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | using clamp, double peak observed in pedestal. Still wrapped in 3 layers teflon, 2 layers black tape. result: 11.6 +- 0.5 pe/MeV | ||
3810 | 10/4/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5031 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 12.5 +- 0.3 pe/MeV | ||
3811 | 10/4/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5031 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 12.4 +- 0.2 pe/MeV | ||
3812 | 10/4/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5036 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape, better end taped shut, open end has bubbles. result: 11.4 +- 0.3 pe/MeV | ||
3813 | 10/4/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5036 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape, better end taped shut, open end has bubbles. result: 10.7 +- 0.1 pe/MeV | ||
3814 | 10/5/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5039 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, equilibrate ~15 min, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 17.4 +- 0.2 pe/MeV | ||
3815 | 10/5/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5039 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 18.0 +- 0.4 pe/MeV | ||
3816 | 10/5/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5034 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, equilibrate ~15 min, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 16.1 +- 0.2 pe/MeV | ||
3817 | 10/5/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5039 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 16.1 +- 0.6 pe/MeV | ||
3818 | 10/5/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5037 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape, note: optical coupling to PMT not very good in this run. result: 23.4 +- 0.6 pe/MeV | ||
3819 | 10/5/2017 | Tanja | XP2262 (SN 34426) | CRYTUR II | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 13.3 +- 0.4 pe/MeV | ||
3820 | 10/5/2017 | Tanja | XP2262 (SN 34426) | CRYTUR II | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 13.8 +- 0.1 pe/MeV | ||
3821 | 10/6/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5033 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 12.4 +- 0.4 pe/MeV | ||
3822 | 10/6/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5033 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 12.2 +- 0.3 pe/MeV | ||
3823 | 10/6/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5040 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, equilibrate ~10min, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 13.8 +- 0.8 pe/MeV | ||
3824 | 10/6/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5040 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 13.5 +- 0.3 pe/MeV | ||
3825 | 10/6/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5035 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 13.5 +- 0.3 pe/MeV | ||
3826 | 10/6/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5035 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 14.1 +- 0.3 pe/MeV | ||
3827 | 10/6/2017 | Ian/Tanja | XP2262 (SN 34426) | J24 (main reference) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | install using clamp, start at about 19 degC, no time to equilibrate, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 15.7 +- 0.2 pe/MeV | ||
3828 | 10/7/2017 | Ian/Tanja | XP2262 (SN 34426) | J24 (main reference) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 15.6 +- 0.1 pe/MeV | ||
3829 | 10/9/2017 | Ian/Tanja | XP2262 (SN 34426) | J24 (main reference) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 15.6 +- 0.1 pe/MeV | ||
3830 | 10/10/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5032 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, equilibrate 30min, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 11.4 +- 0.3 pe/MeV | ||
3831 | 10/10/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5032 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 11.6 +- 0.3 pe/MeV | ||
3832 | 10/10/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5038 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, equilibrate 30min, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 14.8 +- 0.3 pe/MeV | ||
3833 | 10/10/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5038 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: 15.96 +- 0.4 pe/MeV | ||
3834 | 10/10/2017 | Tanja | XP2262 (SN 34426) | SICCAS 5038 (Box3, set 1/160p) | -2.300 (NIM) | 150 (100 effective) | LY at 18°C | after thermal annealing, install using clamp, double peak observed in pedestal. wrapped in 3 layers teflon, 2 layers black tape. result: pe/MeV |