MainPage:Nuclear:NPS:PWO:CrystalLOG:SICCAJ45

From cua_phy
Jump to navigation Jump to search
⇐ Back to Crystal LOG
⇐ Back to Neutral Particle Spectrometer
⇐ Back to the Main_Page

Time LOG

SICCA J45
Date Experimenter(s) Description Comments
05/02/2016 Salina Pictures Documentation of crystal's conditions as it arrived at CUA.
05/02/2016 Marco Thermal annealing Ramp up at 18oC/hour up to 200oC. Stays for 10 hours and ramp down at 18oC/hour.
05/19/2016 Marco Transverse Transmittance At the center of the crystal, every nm from 200nm until 800nm. Sample: 1362
05/20/2016 Marco Transverse Transmittance Scan Wavelengths: 360nm, 420nm, 620nm. Sampling period: 0.6s, moving motor at 2mm/s.
05/25/2016 Salina Light Yield Run 3220: crystal wrapped with ~5 layers of teflon tape and 2 layers of black electrical tape. Steps of 20min, 1cm. T=18°C. LY = 15.38 ± 0.612 pe/MeV.
07/08/2016 Marco Longitudinal Transmittance Scan Wavelengths: 200-800 nm. Sampling period: 0.6s, moving motor at 2mm/s.
07/26/2016 Summer Students Run 3262. 3 layers of teflon tape, 1 layer of electrical tape. ~50,000 incidents
08/09/16 Summer Students 3 layers of teflon 1 later of electrical. ~50,000 incidents at ~18 degrees Celsius. 200ns gatewidth->1000ns gatewidth.

200ns: 3384, 300ns:3385, 400ns:3386, 500ns:3387, 600ns:3388, 700ns:3389, 800ns:3391, 900ns:3393, 1000ns:3289

08/11/2016 Summer Students Light Yield Run 3403. 3 layers of teflon tape, 1 layer of electrical tape. ~50,000 incidents. Light Yield: .

Pictures at 04/29/2016

xxxx.
xxxx.

Light Yield

07/26/2016

Light Yield Gate Width