| 1095 || 12/05/2012 || Marco || Ham. RA1815 || -1.80 (Camac) || Gain || Using plastic dark box, with LED pulse with 7 (on 5V) at 3.5 (on 15ns) with ~1kHz. It is important to note that the PMT pulse has ~30ns and the integration window has 60ns (the maximum from the coincidence module). If the pulse arrives too delayed (~+20ns), part of it may be outside the the integration window. || [[File:1095.png|frameless]]
+
| colspan="8" align="center" | New HV module (VME CAEN V6533N) installed. Scan setup rechecked for bad contacts. Restarting scan project...
 +
|-
 +
| 1224 || 10/01/2013 || Marco || Blast || -1.500 (VME) || Gain || First test to analyze calibration. || [[File:1224.png|frameless]]
 
|}
 
|}

Revision as of 15:00, 1 October 2013

⇐ Back to the PMT Characteristics page
⇒ Go to Summer 2012 work page

LogBook on the CODA data acquisition, at CUA

Page 5, [4], [3], [2], [1].

LogBook - Page 04
Run Number Date User PMT High Voltage Experiment Description and observations Image
New HV module (VME CAEN V6533N) installed. Scan setup rechecked for bad contacts. Restarting scan project...
1224 10/01/2013 Marco Blast -1.500 (VME) Gain First test to analyze calibration.
Error creating thumbnail: Unable to save thumbnail to destination