| 2766 || 11/12/2015 || Marco || XP2262 (SN 34426) || -2.300 (NIM) || Light Yield || Crystal BTCP ###, wrapped with specular reflector (low statistics - 1st test). Na-22 source. T=~13oC (thermometer not in the right position). Mean peak: 353.40 (fit range 170-800).<br>- LY=17.57 pe/MeV (Gain from 2760)<br>- LY=18.56 pe/MeV (Gain from 2763) || [[File:2766.png|frameless]]
 
| 2766 || 11/12/2015 || Marco || XP2262 (SN 34426) || -2.300 (NIM) || Light Yield || Crystal BTCP ###, wrapped with specular reflector (low statistics - 1st test). Na-22 source. T=~13oC (thermometer not in the right position). Mean peak: 353.40 (fit range 170-800).<br>- LY=17.57 pe/MeV (Gain from 2760)<br>- LY=18.56 pe/MeV (Gain from 2763) || [[File:2766.png|frameless]]
 
|-
 
|-
| 2767 || 11/12/2015 || Marco || XP2262 (SN 34426) || -2.300 (NIM) || Light Yield || Same as 2766, high statistics (~900k evts). Peak center: 329.02 (fit range 170-800). <br>'''LY_fit=17.17 pe/MeV''' || [[File:2767.png|frameless]]
+
| 2767 || 11/12/2015 || Marco || XP2262 (SN 34426) || -2.300 (NIM) || Light Yield || Same as 2766, high statistics (~900k evts). Peak center: 329.02 (fit range 170-800). <br>'''LY_fit=18.0 pe/MeV''' || [[File:2767.png|frameless]]
 
|-
 
|-
 
| 2768 || 11/17/2015 || Marco || XP2262 (SN 34426) || -2.300 (NIM) || Light Yield || CRYTUR 10cm sample crystal, wrapped with specular reflector. 18 oC (thermometer positioned). Peak center: (fit range 170-800).<br>'''LY_fit= pe/MeV''' || [[File:2768.png|frameless]]
 
| 2768 || 11/17/2015 || Marco || XP2262 (SN 34426) || -2.300 (NIM) || Light Yield || CRYTUR 10cm sample crystal, wrapped with specular reflector. 18 oC (thermometer positioned). Peak center: (fit range 170-800).<br>'''LY_fit= pe/MeV''' || [[File:2768.png|frameless]]

Revision as of 17:13, 18 November 2015

⇐ Back to the PMT Characteristics page
⇒ Go to Summer 2012 work page

LogBook on the CODA data acquisition, at CUA

Page (last), [6], [5], [4], [3], [2], [1].

LogBook - Last Page
Run Number Date User PMT High Voltage Experiment Description and observations Image
2759 11/12/2015 Marco XP2262 (SN 34426) -2.300 (NIM) Gain LED (green) gain calibration for NPS light yield. Na22 source is also inside, so may have noise. Rate of ~6.5kHz. T=18oC. SEP: 66.67 (fit range 50-85). PED: 27.66 (fit range 25-35).
Error creating thumbnail: Unable to save thumbnail to destination
2760 11/12/2015 Marco XP2262 (SN 34426) -2.300 (NIM) Gain Same as 2759, reduce light intensity. SEP: 63.92 (fit range 50-85). PED: 27.65 (fit range 25-35).
Error creating thumbnail: Unable to save thumbnail to destination
2761 11/12/2015 Marco XP2262 (SN 34426) -2.300 (NIM) Gain Same as 2760; source removed from dark box. SEP: 67.41 (fit range 50-85). PED: 27.63 (fit range 25-35).
Error creating thumbnail: Unable to save thumbnail to destination
2762 11/12/2015 Marco XP2262 (SN 34426) -2.300 (NIM) Gain Same as 2761; slightly higher light intensity. SEP: 67.78 (fit range 50-85). PED: 27.72 (fit range 25-35).
Error creating thumbnail: Unable to save thumbnail to destination
2763 11/12/2015 Marco XP2262 (SN 34426) -2.300 (NIM) Gain Light intensity brought to zero (width=0). SEP: 61.98 (fit range 50-85). PED: 27.63 (fit range 25-35).
Error creating thumbnail: Unable to save thumbnail to destination
2766 11/12/2015 Marco XP2262 (SN 34426) -2.300 (NIM) Light Yield Crystal BTCP ###, wrapped with specular reflector (low statistics - 1st test). Na-22 source. T=~13oC (thermometer not in the right position). Mean peak: 353.40 (fit range 170-800).
- LY=17.57 pe/MeV (Gain from 2760)
- LY=18.56 pe/MeV (Gain from 2763)
Error creating thumbnail: Unable to save thumbnail to destination
2767 11/12/2015 Marco XP2262 (SN 34426) -2.300 (NIM) Light Yield Same as 2766, high statistics (~900k evts). Peak center: 329.02 (fit range 170-800).
LY_fit=18.0 pe/MeV
Error creating thumbnail: Unable to save thumbnail to destination
2768 11/17/2015 Marco XP2262 (SN 34426) -2.300 (NIM) Light Yield CRYTUR 10cm sample crystal, wrapped with specular reflector. 18 oC (thermometer positioned). Peak center: (fit range 170-800).
LY_fit= pe/MeV
Error creating thumbnail: Unable to save thumbnail to destination
2769 11/17/2015 Marco XP2262 (SN 34426) -2.300 (NIM) Light Yield CRYTUR 10cm sample crystal, wrapped with ~5 layers of teflon. Peak center: (fit range 170-800).
LY_fit= pe/MeV
Error creating thumbnail: Unable to save thumbnail to destination