Line 50: Line 50:
 
| 2797..2800 || 12/02/2015 || Marco || XP2262 (SN 34426) || CRYTUR 10cm || -2.300 (NIM) || 100 || LY scan || CRYTUR 10 cm scan, T=18.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm. || [[File:2797.png|frameless]]
 
| 2797..2800 || 12/02/2015 || Marco || XP2262 (SN 34426) || CRYTUR 10cm || -2.300 (NIM) || 100 || LY scan || CRYTUR 10 cm scan, T=18.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm. || [[File:2797.png|frameless]]
 
|-
 
|-
| 2803..2815 || 12/02/2015 || Marco || XP2262 (SN 34426) || CRYTUR 10cm || -2.300 (NIM) || 100 || LY scan || Repeated 2797 scan without CODA crash! CRYTUR 10 cm scan, T=18.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm. || [[File:2803.png|frameless]]
+
| 2803..2815 || 12/02/2015 || Marco || XP2262 (SN 34426) || CRYTUR 10cm || -2.300 (NIM) || 100 || LY scan || Repeated 2797 scan without CODA crash! CRYTUR 10 cm scan, T=18.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm.<br>'''LY=19.77 pe/MeV''' || [[File:2803.png|frameless]]
 
|-
 
|-
| 2821..2832 || 12/03/2015 || Marco || XP2262 (SN 34426) || CRYTUR 10cm || -2.300 (NIM) || 100 || LY scan|| CRYTUR 10 cm scan, T=10.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm, 12 positions. || [[File:2821.png|frameless]]
+
| 2821..2832 || 12/03/2015 || Marco || XP2262 (SN 34426) || CRYTUR 10cm || -2.300 (NIM) || 100 || LY scan|| CRYTUR 10 cm scan, T=10.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm, 12 positions.<br>'''LY=22.99 pe/MeV''' || [[File:2821.png|frameless]]
 
|-
 
|-
 
| 2834 || 12/04/2015 || Marco || XP2262 (SN 34426) || CRYTUR 10cm || -2.300 (NIM) || 36.3 || LY timing || CRYTUR 10 cm, T=10.0 oC. Position center (5cm). (gate delay adjusted to have minimum with no signal inside) || [[File:2834.png|frameless]]
 
| 2834 || 12/04/2015 || Marco || XP2262 (SN 34426) || CRYTUR 10cm || -2.300 (NIM) || 36.3 || LY timing || CRYTUR 10 cm, T=10.0 oC. Position center (5cm). (gate delay adjusted to have minimum with no signal inside) || [[File:2834.png|frameless]]

Revision as of 13:15, 4 December 2015

⇐ Back to the PMT Characteristics page
⇒ Go to Summer 2012 work page

LogBook on the CODA data acquisition, at CUA

Page (last), [6], [5], [4], [3], [2], [1].

LogBook - Last Page
Run Number Date User PMT Crystal High Voltage Gate (ns) Experiment Description and observations Image
2759 11/12/2015 Marco XP2262 (SN 34426) - -2.300 (NIM) 100 Gain LED (green) gain calibration for NPS light yield. Na22 source is also inside, so may have noise. Rate of ~6.5kHz. T=18oC. SEP: 66.67 (fit range 50-85). PED: 27.66 (fit range 25-35).
Error creating thumbnail: Unable to save thumbnail to destination
2760 11/12/2015 Marco XP2262 (SN 34426) - -2.300 (NIM) 100 Gain Same as 2759, reduce light intensity. SEP: 63.92 (fit range 50-85). PED: 27.65 (fit range 25-35).
Error creating thumbnail: Unable to save thumbnail to destination
2761 11/12/2015 Marco XP2262 (SN 34426) - -2.300 (NIM) 100 Gain Same as 2760; source removed from dark box. SEP: 67.41 (fit range 50-85). PED: 27.63 (fit range 25-35).
Error creating thumbnail: Unable to save thumbnail to destination
2762 11/12/2015 Marco XP2262 (SN 34426) - -2.300 (NIM) 100 Gain Same as 2761; slightly higher light intensity. SEP: 67.78 (fit range 50-85). PED: 27.72 (fit range 25-35).
Error creating thumbnail: Unable to save thumbnail to destination
2763 11/12/2015 Marco XP2262 (SN 34426) - -2.300 (NIM) 100 Gain Light intensity brought to zero (width=0). SEP: 61.98 (fit range 50-85). PED: 27.63 (fit range 25-35).
Error creating thumbnail: Unable to save thumbnail to destination
2766 11/12/2015 Marco XP2262 (SN 34426) BTCP ### -2.300 (NIM) 100 Light Yield Crystal BTCP ###, wrapped with specular reflector (low statistics - 1st test). Na-22 source. T=~13oC (thermometer not in the right position). Mean peak: 353.40 (fit range 170-800).
LY=18.0 pe/MeV
Error creating thumbnail: Unable to save thumbnail to destination
2767 11/12/2015 Marco XP2262 (SN 34426) BTCP ### -2.300 (NIM) 100 Light Yield Same as 2766, high statistics (~900k evts). Peak center: 329.02 (fit range 170-800).
LY_fit= pe/MeV
Error creating thumbnail: Unable to save thumbnail to destination
2768 11/17/2015 Marco XP2262 (SN 34426) CRYTUR 10cm -2.300 (NIM) 100 Light Yield CRYTUR 10cm sample crystal, wrapped with specular reflector. 18 oC (thermometer positioned). Peak center: 286.78 (fit range 170-400). SEP: 70.35 (fit range 50-85). PED: 27.36 (fit range 25-35).
LY_fit=11.8 pe/MeV
Error creating thumbnail: Unable to save thumbnail to destination
2769 11/18/2015 Marco XP2262 (SN 34426) CRYTUR 10cm -2.300 (NIM) 100 Light Yield CRYTUR 10cm sample crystal, wrapped with ~5 layers of teflon. T=18oC (2 hours of stabilization only). Peak center: 444.07 (fit range 170-800). SEP and PED: same as 2768 (not enough stats.)
LY_fit= 18.97 pe/MeV
Error creating thumbnail: Unable to save thumbnail to destination
2771 11/19/2015 Marco XP2262 (SN 34426) CRYTUR 10cm -2.300 (NIM) 100 LY Same as 2769, with temperature stable for the whole night; 900k evts.
Error creating thumbnail: Unable to save thumbnail to destination
2774..2779 11/24/2015 Marco XP2262 (SN 34426) CRYTUR 10cm -2.300 (NIM) 100 LY scan CRYTUR 10 cm scan. Same configuration as 2769. Running 10 minutes per position, in steps of 1cm. File:2774.png
2781..2793 11/30/2015 Marco XP2262 (SN 34426) CRYTUR 10cm -2.300 (NIM) 100 LY scan CRYTUR 10 cm scan, temperature NOT stable (just testing CODA and scan auto mode). Same configuration as 2769. Running 10 minutes per position, in steps of 1cm. File:2781.png
2797..2800 12/02/2015 Marco XP2262 (SN 34426) CRYTUR 10cm -2.300 (NIM) 100 LY scan CRYTUR 10 cm scan, T=18.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm. File:2797.png
2803..2815 12/02/2015 Marco XP2262 (SN 34426) CRYTUR 10cm -2.300 (NIM) 100 LY scan Repeated 2797 scan without CODA crash! CRYTUR 10 cm scan, T=18.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm.
LY=19.77 pe/MeV
Error creating thumbnail: Unable to save thumbnail to destination
2821..2832 12/03/2015 Marco XP2262 (SN 34426) CRYTUR 10cm -2.300 (NIM) 100 LY scan CRYTUR 10 cm scan, T=10.0 oC. Same configuration as 2769. Running 90 minutes per position, in steps of 1cm, 12 positions.
LY=22.99 pe/MeV
Error creating thumbnail: Unable to save thumbnail to destination
2834 12/04/2015 Marco XP2262 (SN 34426) CRYTUR 10cm -2.300 (NIM) 36.3 LY timing CRYTUR 10 cm, T=10.0 oC. Position center (5cm). (gate delay adjusted to have minimum with no signal inside) File:2834.png
2835 12/04/2015 Marco XP2262 (SN 34426) CRYTUR 10cm -2.300 (NIM) 46.5 LY timing Same as 2834, adjusted gate width only. File:2835.png
2836 12/04/2015 Marco XP2262 (SN 34426) CRYTUR 10cm -2.300 (NIM) 56.1 LY timing Same as 2834, adjusted gate width only. File:2836.png