Refractive index measurement
File:Refrac index setup.png 1) Experimental setup for the measurement of the aerogel refractive index. A laser sheet is refracted by the tile and hit a target paper. The deviation of the laser can be measured for the calculation of the refractive index. |
File:4491.png 2) Part of the laser sheet pass through the aerogel tile, being deflected by its refractive index. The other part pass on the top of the tile, going direct to the reference paper target. |
|
Kaon Aerogel Detector prototype - Aerogel light yield comparison
In order to compare the different aerogel (refractive index) we will use in the Kaon Aerogel Detector, a set of cosmic rays data was acquired with the CUA's prototype. Details of the prototype can be found in the Prototype page.
To evaluate the relative light yield of the aerogel tiles with refractive indexes 1.030 and 1.015, the setup was run with the same configuration twice (just exchanging the aerogel).
The setup configuration was kept as follows:
- Cosmic ray configuration (two triggering PMTs with scintillator on the top and bottom of the prototype)
- 08 aerogel tiles
- Mylar covering the internal walls of the prototype
- High voltage of -1.80 kV on the PMT
- Blast PMT for all runs (model XP4500/B s/n 09650, with calibration from run 481)
To see a similar test done for the analysis of the different PMTs options for the Kaon Aerogel Detector, click here.