Run Number | Date | User | PMT | High Voltage | Experiment | Description and observations | Image |
---|---|---|---|---|---|---|---|
1095 | 12/05/2012 | Marco | (Hamamatsu) | -1.80 (Camac) | Gain | Using plastic dark box, with LED pulse with 7 (on 5V) at 3.5 (on 15ns) with ~1kHz. It is important to note that the PMT pulse has ~30ns and the integration window has 60ns (the maximum from the coincidence module). If the pulse arrives too delayed (~+20ns), part of it may be outside the the integration window. | File:1095.png |
1096 | 12/05/2012 | Marco | (Hamamatsu) | -1.80 (Camac) | Gain | Repeating run #1095 with the minimum light that still give some pulses once in a while. | Error creating thumbnail: Unable to save thumbnail to destination
|